FDA Medical devices recall ·

DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model recall

Recall number
Z-2861-2026
Date
(initiated June 25, 2025)
Company / brand
Nuclein LLC
8305 Cross Park Dr, Austin, TX, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
2
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: DASH Rapid PCR Instrument Product Name: DASH Rapid PCR Instrument Model/Catalog Number: SG-0002 Software Version: 6.0.2 Product Description: The DASH¿ Rapid PCR Instrument (DASH Instrument) is an automated, fluorescence-based, reverse-transcription polymerase chain reaction (RT-PCR) instrument that extracts specimens and performs nucleic acid amplified tests (NAAT) in DASH test cartridges for the qualitative detection of infectious diseases. The DASH Instrument is intended for use in conjunction with DASH Rapid PCR System tests to aid in the diagnosis of infectious diseases. Component: No.

Why was it recalled?

Due to a software issue which may result in false co-infection (e.g., Positive for Flu A, when SARS-CoV-2 is present).

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: SN: C01250010; UDI: (01)00850063063005(11)250218(21)C01250010; SW: 6.0.2 SN: C02250006; UDI:(01)00850063063005(11)250225(21)C02250006; SW: 6.0.2

Distribution

US Nationwide distribution in the state of Texas.

States: Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2861-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.