FDA Medical devices recall ·
OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model recall
- Recall number
- Z-2857-2026
- Date
- (initiated June 11, 2026)
- Company / brand
- Oxos Medical Inc
1100 Peachtree St Ne Ste 700, Atlanta, GA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 77
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: OXOS Medical, Inc Product Name: MC2 Portable X-Ray System Model/Catalog Number: MC2, MC2.1 Software Version: 5.7.0 Product Description: The MC2 Portable X-Ray System ( MC2 System or MC2 ) is a portable and handheld x-ray system designed to aid clinicians with point of care visualization through diagnostic x-rays of extremities and shoulders. The device allows a clinician to select desired technique factors best suited for their patient anatomy. The MC2 consists of two major system components: the emitter and the cassette. The MC2 emitter and cassette are battery-powered and can be charged with either a wired or wireless charger. The system is intended to interface wirelessly to an external tablet or display when used with the OXOS Device App. The MC2 utilizes an infrared Tracking System and Viewfinder that allows the operator to position the emitter within a safe source-to-image distance (SID) and allows the operator to visualize the anatomy and x-ray field projection prior to initiating an x-ray acquisition. The MC2 also utilizes a LIDAR system to ensure patient safety by maintaining a safe source-to-skin distance. Component: The MC2 Portable X-Ray System is not a component.
Why was it recalled?
Due to a defective circuit board that overheats
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Lot Code: MC2; UDI 00860008892782; Serial Numbers 1523, 1527, 1488, 1528, 1529, 1544 MC2.1; UDI 00850079606043; Serial Numbers: 1461, 1462, 1547 Full UDI for affected devices: (01)00860008892782(11)260107(21)1523 (01)00860008892782(11)260129(21)1527 (01)00860008892782(11)260109(21)1488 (01)00860008892782(11)260123(21)1528 (01)00860008892782(11)260130(21)1529 (01)00860008892782(11)260227(21)1544 (01)00850079606043(11)260407(21)1461 (01)00850079606043(11)260407(21)1462 (01)00850079606043(11)260408(21)1547
Distribution
US Nationwide distribution in the states of FL, GA, IN, SC, TX, VA.