FDA Medical devices recall ·

GEM Premier 5000 with IQM2 Model recall

Recall number
Z-2821-2026
Date
(initiated May 8, 2026)
Company / brand
Instrumentation Laboratory
180 Hartwell Rd, Bedford, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
122
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

GEM Premier 5000 with IQM2 Model/Catalog Number: 00024019255 and GEM Premier 7000 with iQM3 Model/Catalog Number: 00000015279 Portable blood analyzer system. Affected Accessory: Bar Code Reader (BCR) Wand Part No. 00024015859

Why was it recalled?

Potential for Bar Code Reader (BCR) wands, used with the GEM Premier 5000 and GEM Premier 7000 systems, to assign patient results to an incorrect patient ID.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

GEM Premier 5000 Part number: 00024019255, UDI: 08426950870227 GEM Premier 7000 Part number: 00000015279, UDI: 0195226000924 Affected Accessory: Bar Code Reader (BCR) Wand that received a configuration update on or after 8 April 2026. Part No. 00024015859; UDI No. 08426950869634

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2821-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.