FDA Medical devices recall ·

ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and… recall

Recall number
Z-2723-2026
Date
(initiated June 10, 2026)
Company / brand
ICU Medical, Inc.
600 N Field Dr, Lake Forest, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
12525 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ICU Medical Plum Duo Infusion Pump, v1.2.3: 1) Item Number 400020401 and 2) Item Number 400020403; infusion pump

Why was it recalled?

In a specific scenario, the pump will not decrement the Volume to be Infused (VTBI) displayed on the Delivery screen and may deliver up to twice the VTBI at the programmed rate.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1) UDI/DI +M335400021, Serial Numbers: 50003704, 50003710, 50004046, 50004072, 50004149, 50004154, 50004156, 50004163, 50004166, 50004168, 50004177, 50004183, 50004205, 50004218, 50004220, 50004249, 50004258, 50004262, 50004263, 50004264, 50004271, 50004300, 50004303, 50004304, 50004305, 50004307, 50004308, 50004310, 50004312, 50004314, 50004316, 50004318, 50004319, 50004323, 50004324, 50004325, 50004326, 50004328, 50004330, 50004331, 50004332, 50004333, 50004334, 50004335, 50004336, 50004338, 50004339, 50004340, 50004341, 50004342, 50004344, 50004345, 50004346, 50004347, 50004348, 50004349, 50004350, 50004351, 50004352, 50004356, 50004358, 50004359, 50004361, 50004363, 50004365, 50004366, 50004370, 50004372, 50004373, 50004374, 50004376, 50004380, 50004382, 50004383, 50004384, 50004385, 50004386, 50004387, 50004389, 50004390, 50004391, 50004392, 50004393, 50004394, 50004395, 50004396, 50004398, 50004399, 50004401, 50004403, 50004405, 50004407, 50004409, 50004410, 50004411, 50004413, 50004415, 50004416, 50004417, 50004424, 50004425, 50004426, 50004428, 50004431, 50004433, 50004435, 50004436, 50004438, 50004439, 50004442, 50004443, 50004444, 50004445, 50004448, 50004449, 50004451, 50004452, 50004455, 50004458, 50004460, 50004461, 50004462, 50004464, 50004465, 50004467, 50004468, 50004469, 50004471, 50004472, 50004473, 50004474, 50004477, 50004481, 50004482, 50004486, 50004487, 50004493, 50004495, 50004499, 50004501, 50004502, 50004506, 50004510, 50004512, 50004513, 50004515, 5

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2723-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.