FDA Medical devices recall ·

ABIOMED Impella CP Smart Assist Set, JP, Pump Set Code 0048-0024-JP recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2690-2026
Date
(initiated June 11, 2026)
Company / brand
Abiomed, Inc.
22 Cherry Hill Dr, Danvers, MA, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
8,382 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ABIOMED Impella CP Smart Assist Set, JP, Pump Set Code 0048-0024-JP; containing Introducer Product Code 0052-3025-JP. heart pump and accessories

Why was it recalled?

A potential for introducer sheath leakage from under the sheath cap and along the hub score lines, identified in 14Fr and 23Fr Introducers. These manufacturing issues are not visually detectable by the user, and 14Fr or 23Fr Introducers are provided in every pump set.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Pump Set Code 0048-0024-JP; Introducer Product Code 0052-3025-JP. GTIN: 00813502011609. All current batches in field.

Distribution

Worldwide distribution. US Nationwide, Australia (AU), Austria (AT), Belgium (BE), Brazil (BR), Brunei (BN), Canada (CA), China (CN), Croatia (HR), Czech Republic (CZ), Denmark (DK), Finland (FI), France (FR), Germany (DE), Greece (GR), Hong Kong (HK), Ireland (IE), Israel (IL), Italy (IT), Japan (JP), Kuwait (KW), Luxembourg (LU), Malaysia (MY), Mexico (MX), Netherlands (NL), Norway (NO), Panama (PA), Poland (PL), Saudi Arabia (SA), Serbia (RS), Slovenia (SI), South Korea (KR), Spain (ES), Sweden (SE), Switzerland (CH), Taiwan (TW), Thailand (TH), United Arab Emirates (AE), and United Kingdom (GB).

States: California Illinois

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2690-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.