FDA Medical devices recall ·

ZAP-X Radiosurgery System, Model: 300150 recall

Recall number
Z-2639-2025
Date
(initiated July 25, 2025)
Company / brand
Zap Surgical Systems
590 Taylor Way, San Carlos, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
17
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ZAP-X Radiosurgery System, Model: 300150

Why was it recalled?

If the radiosurgery system triggers a proximity error message during a long gantry move (greater-than180 degrees), and a subsequent proximity error message occurs after the automatic reduction of speed, a software defect could potentially lead the collimator to collide with patient shoulders or the patient table.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI: 00860183001504, Serial Numbers: ZUC17153, ZUC20156, ZUC20157, ZUC20159, ZUC21162, ZUC21164, ZUC22166, ZUC22169, ZUC22170, ZUC22171, ZUC22173, ZUC22174, ZUC23176, ZUC23177, ZUC24179, ZUC24180, ZUC24181. Software version: TDS V1.10.1

Distribution

Worldwide - US Nationwide distribution in the states of FL, CO, NJ, NY and the countries of Germany, France, Spain, Poland, Republic of Korea, T¿rkiye, Paraguay.

States: Colorado Florida New Jersey New York

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2639-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.