FDA Medical devices recall ·

Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No recall

Recall number
Z-2636-2025
Date
(initiated August 6, 2025)
Company / brand
Maquet Cardiovascular, LLC
45 Barbour Pond Dr, Wayne, NJ, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
98 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Vasoview Hemopro 3 Endoscopic Vessel Harvesting System Model No. C-VH-6000. The VASOVIEW HEMOPRO 3 Endoscopic Vessel Harvesting (EVH) System (VH-6000, VH-6001, and VH-6010) is a vessel harvesting kit designed for use with a 7 mm endoscope to cut and cauterize vessel branches. The HP3 Harvesting Tool features two curved, insulated jaws with heating elements that operate through heat and pressure rather than radiofrequency energy.

Why was it recalled?

Reports of out-of-box failures discovered during the limited launch phase. Failure modes included angled deformity at the attachment site of the Harvesting Tool Jaws and the distal end of the Harvesting Tool Shaft, jaws not fully closed out of box, and heater wire lifted/flexed away from jaws.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model No. C-VH-6000; UDI: 00607567701274; Lot/Serial No. 3000486429, 3000487231, 3000487726, 3000488639.

Distribution

US Nationwide distribution in the states of FL, IL, MD & TX.

States: Florida Illinois Maryland Texas

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2636-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.