FDA Medical devices recall ·
Pinnacle Radiation Therapy Planning System: Multimodality Simulation… recall
- Recall number
- Z-2570-2025
- Date
- (initiated August 5, 2025)
- Company / brand
- Philips Medical Systems (Cleveland) Inc
3630 Sw 47th Ave, Gainesville, FL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 810 (350 U.S. and 460 O.U.S.)
- Hazard tags
- Software
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Pinnacle Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Pinnacle TumorLOC 16.2 Model 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Pinnacle TumorLOC 16.2.1 Model 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Pinnacle TumorLOC 16.0.2 Model 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Pinnacle TumorLOC 14.0 Model 870200 UDI code: N/A Pinnacle3 Radiation Therapy Planning System is a software package intended to provide planning support for the treatment of disease processes. The full Pinnacle3 Radiation Therapy Planning System software package provides planning support for the treatment of disease processes, utilizing photon, electron and brachytherapy techniques.
Why was it recalled?
Due to a software issue, there is a potential image error of the Region of Interest for expansion/contraction for HFP (Head First Prone), FFS (Feet First Supine) and FFP (Feet First Prone) orientations.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Pinnacle¿ Radiation Therapy Planning System: Multimodality Simulation Workspace 18.0.5 Model Number: 870258 UDI code: (01)00884838103566(10)18.0.5.22236 Lot Number (Software Version): 18.0.5 Pinnacle TumorLOC 16.2 Model Number: 870226 UDI code: (01)00884838091122(10)16.2.0.60010 Lot Number (Software Version): 16.2 Pinnacle TumorLOC 16.2.1 Model Number: 870227 UDI code: (01)00884838091122(10)16.2.1.60098 Lot Number (Software Version): 16.2.1 Pinnacle TumorLOC 16.0.2 Model Number: 870218 UDI code: (01)00884838085602(10)16.0.2.60004 Lot Number (Software Version): 16.0.2 Pinnacle TumorLOC 14.0 Model Number: 870200 UDI code: N/A Lot Number (Software Version): 14.0
Distribution
U.S Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV.
States: Alabama Arkansas Arizona California Colorado Connecticut Delaware Florida Georgia Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2570-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.