FDA Medical devices recall ·

Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCK use… recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2539-2025
Date
(initiated August 12, 2025)
Company / brand
Max Mobility LLC
300 Duke Dr, Lebanon, TN, United States
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
15,834 units
Hazard tags
Injury risk
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Max Mobility/Permobil Smart Drive MX2+ SpeedControl Dial Ref: MX2-3DCK use in conjunction with the SmartDrive MX2+ Wheelchair Power Assist component.

Why was it recalled?

Due to a faulty electrical connection between the speed control Dial and the wheelchair motor, their is a potential for lost of control of the device potentially resulting in minor or major injuries.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

All Serial Numbers distributed between the dates of April 25, 2022 - July 8, 2025

Distribution

US: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD ME MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY OUS: CAN, NZL, KOR, BMU, GBR, JPN, TWN, AUS, BRA, COL, MEX, ABW, SWE, CHL, ARG, HKG

States: Alaska Arkansas Arizona California Connecticut District of Columbia Florida Georgia Iowa Illinois Kansas Kentucky Maryland Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2539-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.