FDA Medical devices recall ·

Merge Hemo, Model RCSV2, Model recall

Recall number
Z-2518-2025
Date
(initiated August 11, 2025)
Company / brand
Merge Healthcare, Inc.
900 Walnut Ridge Dr, Hartland, WI, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
84
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Merge Hemo, Model RCSV2, Model/Catalog Number 99-01172-00, software versions 11.1 and 11.1.1 Product Description: The Merge Hemo, Model RCSV2 device is a hemodynamics recording computer system that monitors, measures, displays, records and stores various physiologic and blood flow parameters based on the output from one or more electrodes, transducers, or measuring devices. The system is used for documenting and monitoring patients undergoing diagnostic and interventional catheter-based procedures. The procedures can be performed in the clinical areas of invasive cardiology, interventional radiology, and cardiac electrophysiology.

Why was it recalled?

Merge Healthcare has been informed by Schiller AG (Baar, Switzerland) of an issue with the Schiller ARGUS PB-3000 Patient Data Module (PDM) where certain units trigger an error message on the Merge Hemo Client ( Pressure Sensor Defective ) during non-invasive blood pressure (NIBP) measurements, specifically when the initial inflation pressure is set in the high-pressure range (>240 mmHg). The ARGUS PB-3000 is a vital signs monitoring instrument intended to be integrated into the Merge Hemo System. No other components of the Merge Hemo System are affected.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Schiller Model Name: ARGUS PB-3000, Schiller Model #: 3.921002, Affected PB-3000 Serial Numbers: 7012.000035, 7012.000036, 7012.000037, 7012.000038, 7012.000041, 7012.000042, 7012.000044, 7012.000045, 7012.000046, 7012.000047, 7012.000048, 7012.000050, 7012.000051, 7012.000052, 7012.000053, 7012.000057, 7012.000058, 7012.000059, 7012.000060, 7012.000061, 7012.000062, 7012.000063, 7012.000064, 7012.000065, 7012.000066, 7012.000067, 7012.000068, 7012.000069, 7012.000071, 7012.000072, 7012.000073, 7012.000074, 7012.000075, 7012.000076, 7012.000078, 7012.000079, 7012.000080, 7012.000081, 7012.000082, 7012.000083, 7012.000085, 7012.000087, 7012.000088, 7012.000089, 7012.000090, 7012.000091, 7012.000092, 7012.000093, 7012.000094, 7012.000097, 7012.000098, 7012.000099, 7012.000100, 7012.000101, 7012.000102, 7012.000103, 7012.000104, 7012.000105, 7012.000106, 7012.000107, 7012.000108, 7012.000109, 7012.000110, 7012.000111, 7012.000112, 7012.000113, 7012.000114, 7012.000115, 7012.000116, 7012.000117, 7012.000118, 7012.000119, 7012.000120, 7012.000123, 7012.000124, 7012.000125, 7012.000126, 7012.000137, 7012.000138, 7012.000140, 7012.000141, 7012.000142, 7012.000157, 7012.000176

Distribution

US Nationwide distribution including in the states of Arizona, Arkansas, California, Florida, Louisiana, Mississippi, North Carolina, North Dakota, Oklahoma, Pennsylvania, Georgia, Illinois, Kentucky, Tennessee, Texas, Washington, and Wisconsin.

States: Arkansas Arizona California Florida Georgia Illinois Kentucky Louisiana Mississippi North Carolina North Dakota Oklahoma Pennsylvania Tennessee Texas Washington Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2518-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.