FDA Medical devices recall ·
RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X1100… recall
- Recall number
- Z-2486-2025
- Date
- (initiated July 22, 2025)
- Company / brand
- Reflexion Medical, Inc.
25881 Industrial Blvd, Hayward, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 11
- Hazard tags
- Software
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
RefleXion X1 Model RXM1000 UDI Codes: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012 The RefleXion X1 is a radiation therapy system that combines the functionality of an emission computed tomography detection system and a linear accelerator. The device is intended for use with approved fludeoxyglucose F18. The emission computed tomography detection system acquires images of positron-emitting fludeoxyglucose F18 for the purpose of guiding the delivery of megavoltage X-rays for oncologic treatment with radiation therapy using an FDA cleared, authorized, or approved linear accelerator.
Why was it recalled?
Due to incorrect software configuration that potentially allows more than two (2) fractions within a 12-hour period (fractionLimits) and could potentially lead to adverse events (toxicity).
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Model Number: RXM1000 UDI Codes that includes Serial Numbers: (01)00860003983812(11)230504(21)X11001 (01)00860003983812(11)230518(21)X11002 (01)00860003983812(11)230505(21)X11003 (01)00860003983812(11)230530(21)X11004 (01)00860003983812(11)230811(21)X11006 (01)00860003983812(11)230410(21)X11007 (01)00860003983812(11)230803(21)X11008 (01)00860003983812(11)240528(21)X11009 (01)00860003983812(11)240820(21)X11010 (01)00860003983812(11)241107(21)X11011 (01)00860003983812(11)241206(21)X11012
Distribution
U.S. Nationwide distribution in the states of CA, CT, LA, NJ, OR, PA, and TX.
States: California Connecticut Louisiana New Jersey Oregon Pennsylvania Texas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2486-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.