Breathing circuit set, coaxial recall

This is an FDA Class I recall — the most serious type.

Recall number
Z-2463-2026
Date
(initiated May 29, 2026)
Company / brand
Hamilton Medical AG
Parc Industrial Vial 10, Domat/Ems, N/A, Switzerland
Category
Medical devices › Class I
Classification
Class I
Status
Ongoing
Units affected
27,270 pcs
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Breathing circuit set, coaxial: REF: 260127, 260128, 260167, 260168 used with HAMILTON-C1, HAMILTON-T1 and HAMILTON-MR1 ventilators.

Why was it recalled?

Ventilator coaxial breathing circuit may have, expiratory valve sets with membranes sticking to valve body, not detectable during pre-operative test, which will result in "exhalation obstructed " alarm, which may result in inadequate ventilation, oxygen desaturation. Advising: Prior to use, test lung action. If alarm occurs, expiratory valve set removal to overcome membrane adhesion or replacement

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

REF/UDI-DI: 260127/07630002802956, 17630002802953; 260128/07630002802963, 17630002802960; 260167/07630002802970, 17630002802977; 260168/07630002802987, 17630002802984. Lot numbers between 200379 and 205050 (including both.)

Distribution

US Nationwide distribution including in the states of KS, WA, VT, TX, WI, AR, OH, PA, CA, CO, ID, AZ, ME, MO, MI, IL, OR, VA, FL, NJ, SD, NY, NM, TN, CT, NE, OK, AL, GA, KY, NH, WY, ND, IN, UT, WV, DE, IA, PR, MN, LA, AK, HI.

States: Alaska Alabama Arkansas Arizona California Colorado Connecticut Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Maine Michigan Minnesota Missouri North Dakota Nebraska New Hampshire New Jersey New Mexico New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming

What does Class I mean?

Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.

Source: FDA (medical devices), recall Z-2463-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.