FDA Medical devices recall ·
Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel… recall
- Recall number
- Z-2462-2025
- Date
- (initiated June 18, 2025)
- Company / brand
- C.R. Bard Inc
8195 Industrial Blvd Ne, Covington, GA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 175,158
- Hazard tags
- Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Bardex Lubri-Sil I.C. 400-series Temperature Sensing Silver/Hydrogel Coated All-Silicone Foley Catheter 5cc Balloon, REF: 119314; SureStep Foley Tray System Lubri-Sil I.C. Complete Care, REF: A319514A; Bard Silver Lubri-Sil Foley Tray Temperature Sensing with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29000J14, 29030J14, 29000J14S; Bard Silver Lubri-Sil Foley Tray Temperature Sensing, Complete Care with Bard Hydrogel and Bacti-Guard Silver Alloy Coating, 14Fr, REF: 29030J14S
Why was it recalled?
Foley catheter may have an obstructed lumen, which may result in an inability to drain the bladder, of urine and may include pelvic or abdominal pain, flank pain, bladder dysfunction, obstructive uropathy, acute kidney injury and/or hydronephrosis.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Catalog/UDI-DI/Lot(Expiration): 119314/00801741039843, 10801741039840/NGJS3588(31/03/2029), NGKQ2500(21/12/2029), NGKQ0170(30/11/2029), NGJW2499(31/08/2029), NGJT1284 (31/03/2029), NGKQ2492(31/12/2029), NGKP3152(31/12/2029), NGJX0215(31/07/2029). A319514A/00801741100321, 10801741100328/NGJN0498(31/01/2026), NGJW3127(30/04/2027) 29000J14/21200BZY00110000/MYJX0683(28/05/2027), MYJP0751(28/02/2027), MYJR1984(28/04/2027), MYJQ0070(28/04/2027), MYJS6278(28/04/2027), MYJR3026(28/04/2027), MYJT3766(28/04/2027), MYJY6197(28/06/2027), MYJX2833(28/06/2027), MYJQ4022(28/04/2027), MYJT3764(28/04/2027), MYJZ0962(28/06/2027), MYJS4679(28/04/2027), MYJX0686(28/06/2027), MYJV4359(28/03/2027), MYJS3597(28/04/2027), MYJW1900(28/04/2027), MYJT3765(28/04/2027), MYJY1245(28/06/2027), MYJY1242(28/06/2027), MYJY1248(28/06/2027). 29030J14/21200BZY00110000/MYJX2846(28/06/2027), MYJP2013(28/02/2027), MYJT1017(28/04/2027), MYJW1906(28/04/2027), MYJT4535(28/07/2027), MYJP4407(28/04/2027), MYJS1895(28/04/2027), MYJW1904(28/04/2027), MYJV4102(28/03/2027), MYJX2836(28/05/2027), MYJZ4526(28/06/2027), MYJT3468(28/04/2027), MYJP2012(28/02/2027), MYJV4099(28/03/2027), MYJZ0964(28/06/2027), MYJQ0071(28/04/2027), MYJS6279(28/04/2027), MYJY4511(28/06/2027), MYJQ4024(28/04/2027), MYJZ0963(28/06/2027), MYJT1018(28/04/2027), MYJR3027(28/04/2027), MYJS3599(28/04/2027), MYJQ4023(28/04/2027), MYJR3083(28/04/2027), MYJY4512(28/06/2027), MYJQ0073(28/04/2027), MYJS1893(28/04/2027), MYJN2580(28/01/2027), MYJU2705(28/08/202
Distribution
Worldwide - US Nationwide distribution in the states of IA, PA, AR, OR, MN, IN, WV, OK, AL, MD, HI, GA, AZ, MA, FL, CA and the countries of Japan.
States: Alabama Arkansas Arizona California Florida Georgia Hawaii Iowa Indiana Massachusetts Maryland Minnesota Oklahoma Oregon Pennsylvania West Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2462-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.