FDA Medical devices recall ·
Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model recall
- Recall number
- Z-2422-2025
- Date
- (initiated July 24, 2025)
- Company / brand
- Dentsply IH, Inc.
590 Lincoln St, Waltham, MA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 780 units
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: Atlantis IO FLO Product Name: Atlantis IO FLO-S Kit Model/Catalog Number: A04S / 6802034 Product Description: The Atlantis intraoral feature locating object for suprastructures (IO FLO-S) is intended to support the prosthetic procedure with the aim of restoring chewing function in partial or complete edentulous patients, by enabling digital impression taking on implant or abutment level for all positions in the mouth of the patient or on the master model. Component: not a component
Why was it recalled?
The defect in the thread area will not allow the device to fully engage with an implant or analog (implant replica used for prosthetic modeling). The threads will only engage approximately 2 rotations as opposed to full engagement which is approximately 5 rotations. This equates to approximately 1mm difference in depth of engagement.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Catalog / Ref #: 68020034; UDI: 07392532249440; Lot #s: 10515863, 10512084, 10505543, 10556741;
Distribution
Worldwide - US Nationwide distribution in the states of IL, LA, MA, MN, MO, ND, NJ, NM, OH, OK, PA, TX, WV and the countries of AT, AU, BE, CA, DE, DK, ES, FI, FR, IT, JP, NL, NO, PL, QA, SE, UA, UK.
States: California Delaware Illinois Louisiana Massachusetts Minnesota Missouri North Dakota New Jersey New Mexico Ohio Oklahoma Pennsylvania Texas West Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2422-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.