FDA Medical devices recall ·

EXENT Analyser Product Name: EXENT Analyser Model recall

Recall number
Z-2362-2026
Date
(initiated April 16, 2026)
Company / brand
The Binding Site Group, Ltd.
8 Calthorpe Road, Birmingham, N/A, United Kingdom
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: EXENT Analyser Product Name: EXENT Analyser Model/Catalog Number: IE800.A Software Version: 1.0.20 Product Description: The EXENT Analyser is an automated analyser intended for the qualitative and quantitative in vitro measurement of analytes in human body fluids used in conjunction with the EXENT assays.

Why was it recalled?

It was reported that plates 2 and 3 were marked as passed , although it appeared that QC had not been performed on these plates, despite samples being measured. The potential clinical hazards include false or missed detection of M-protein (false positive); and falsely elevated or decreased M-protein concentration.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: Model No; IE800A; EXENT Analyser; UDI; 05051700020886, Serial Number; 101150, Software version; v1.0.20.

Distribution

US Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2362-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.