FDA Medical devices recall ·

HemosIL LMW Heparin Controls recall

Recall number
Z-2323-2025
Date
(initiated February 3, 2025)
Company / brand
Instrumentation Laboratory
180 Hartwell Rd, Bedford, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
41,824 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

HemosIL LMW Heparin Controls; Part Number: 0020300200;

Why was it recalled?

Multiple complaints indicating lower than expected quality control (QC) results and in some instances QC recovery below the lower limit of the package insert acceptance range when testing is performed with HemosIL Liquid Anti-Xa and HemosIL Heparin Calibrator. In some complaints, failure to pass QC resulted in patient testing delays.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Part Number: 0020300200; UDI-DI: 08426950472490; Lot Numbers: N0330228, N0632773, N0935135, N1036592, N0148986, N0441476, N0643673, N0845161, N1147397;

Distribution

Worldwide - US Nationwide and the countries of Algeria, Andorra, Argentina, Australia, Austria, Belarus, Brazil, Brunei, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Denmark, Finland, Germany, Gibraltar, Greece, Guatemala, Hong Kong, Hungary, India, Ireland, Israel, Italy, Japan, Korea, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Macao, Malaysia, Mexico, Panama, Paraguay, Peru, Poland, Portugal, Qatar, Romania, Russia, Saudi Arabia, Serbia, South Africa, Spain, Sweden, Switzerland, Thailand, Tunisia, Turkey, United Kingdom, Uruguay, Vietnam.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2323-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.