FDA Medical devices recall ·
Allia IGS 3 Pulse angiographic X-ray system recall
- Recall number
- Z-2314-2026
- Date
- (initiated April 20, 2026)
- Company / brand
- GE Medical Systems, LLC
3000 N Grandview Blvd, Waukesha, WI, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 9 units
- Hazard tags
- Electrical shock
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Allia IGS 3 Pulse angiographic X-ray system
Why was it recalled?
GE HealthCare has become aware of a potential issue affecting an electrical cable on certain IGS systems equipped with Gaia generators. This issue could lead to a risk of electrical shock to service personnel performing installation, maintenance, or troubleshooting activities inside the C-FRT cabinet while power is ON.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 00195278719577; Serial Numbers: M3-24-053, M3-24-124, M3-25-029, M3-25-051, M3-25-053, M3-25-091, M3-25-098, M3-25-124, M3-26-010, M3-26-030
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2314-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.