FDA Medical devices recall ·

High Pressure Tubing recall

Recall number
Z-2293-2025
Date
(initiated June 5, 2025)
Company / brand
Merit Medical Systems, Inc.
1600 W Merit Pkwy, South Jordan, UT, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
8040
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

High Pressure Tubing: REF: 502211002, HP9720E/C, HP9721E/C

Why was it recalled?

High pressure tubing may relax over time causing it to partially separate from the connector resulting in a void that can trap air in the line. Use of the affected product may result in the trapped air being introduced into the patient.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

REF/UDI-DI/Lot(Expiration): 502211002/I3116954(8/16/2028); HP9720E/C/00884450859889/I3116059(12/10/2027), I3127321(12/30/2027), I3137220(1/28/2028); HP9721E/C/00884450859896/I3116060(12/10/2027)

Distribution

US Nationwide distribution in the states of OH, CA, SD, VA, AZ, MT, TX, MI, FL, TN, PA, NJ, MN, ND, IA, LA, UT, KY, IL, OK, ME, IN, NY, NM, SC, AL, ID, CT, WI, GA, MA, NC, AR, WA.

States: Alabama Arkansas Arizona California Connecticut Florida Georgia Iowa Idaho Illinois Indiana Kentucky Louisiana Massachusetts Maine Michigan Minnesota Montana North Carolina North Dakota New Jersey New Mexico New York Ohio Oklahoma Pennsylvania South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2293-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.