FDA Medical devices recall ·

Azurion system recall

Recall number
Z-2227-2026
Date
(initiated April 28, 2026)
Company / brand
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1651 units (592 US, 1359 OUS)
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Azurion system; System Code Description (Model Numbers): Azurion 3 M12 (722063), Azurion 3 M15 (722064), Azurion 7 B12 (722067), Azurion 7 B20 (722068), Azurion 7 M12 (722078), Azurion 7 M20 (722079, 722224).

Why was it recalled?

The deaeration hose in X-ray tube cooling units type CU3101, manufactured between February 2016 and May 2020, may degrade over time, potentially resulting in oil leakage. Oil leakage may affect the cooling performance of the X-ray tube. When the oil flow drops below a certain threshold, the system automatically switches to low-dose fluoroscopy and displays the message Low load fluoroscopy flavor selected: Tube cooler problem . This condition cannot be resolved by restarting the system.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. Azurion 3 M12; Model: 722063; UDI: (01)00884838085275(21)64, (01)00884838085275(21)56, (01)00884838085275(21)77, (01)00884838085275(21)82, (01)00884838085275(21)47, (01)00884838085275(21)80, (01)00884838085275(21)102, (01)00884838085275(21)117, (01)00884838085275(21)2, (01)00884838085275(21)99, (01)00884838085275(21)44, (01)00884838085275(21)3, (01)00884838085275(21)62, (01)00884838085275(21)61, (01)00884838085275(21)22, (01)00884838085275(21)29, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)30, (01)00884838085275(21)25, (01)00884838085275(21)19, (01)00884838085275(21)18, (01)00884838085275(21)17, (01)00884838085275(21)16, (01)00884838085275(21)14, (01)00884838085275(21)53, (01)00884838085275(21)103, (01)00884838085275(21)101, (01)00884838085275(21)41, (01)00884838085275(21)104, (01)00884838085275(21)93, (01)00884838085275(21)115, (01)00884838085275(21)73, (01)00884838085275(21)128, (01)00884838085275(21)4, (01)00884838085275(21)50, (01)00884838085275(21)72, (01)00884838085275(21)38, (01)00884838085275(21)21, (01)00884838085275(21)78, (01)00884838085275(21)23, (01)00884838085275(21)76, (01)00884838085275(21)63, (01)00884838085275(21)69, (01)00884838085275(21)59, (01)00884838085275(21)51, (01)00884838085275(21)92, (01)00884838085275(21)75, (01)00884838085275(21)94, (01)00884838085275(21)100, (01)00884838085275(21)54, (01)00884838085275(21)46, (01)00884838085275(21)26, (01)00884838085275(21)71, (01)00884838085275(21)106, (01)00884838085275(21)83,

Distribution

Domestic: Nationwide; International: Canada, Algeria, Argentina, Australia, Austria, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Brunei Darussalam, Bulgaria, Chile, China, Colombia, Costa Rica, Croatia, Cuba, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, Ethiopia, Finland, France, French Polynesia, Georgia, Germany, Greece, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Korea, Republic of, Kosovo, Kuwait, Kyrgyzstan, Lao People's Democratic Republic, Latvia, Lebanon, Libya, Lithuania, Malaysia, Maldives, Malta, Martinique, Mexico, Moldova, Mongolia, Morocco, Myanmar, Nepal, Netherlands, New Caledonia, New Zealand, Nigeria, Norway, Oman, Pakistan, Palestine, State of, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, R¿union, Romania, Russian Federation, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, Sri Lanka, Sudan, Sweden, Switzerland, Taiwan, Thailand, Trinidad and Tobago, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Viet Nam, Yemen;

States: Georgia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2227-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.