FDA Medical devices recall ·

Philips Azurion systems not configured with an optional auxiliary pan… recall

Recall number
Z-2225-2026
Date
(initiated April 28, 2026)
Company / brand
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
5,537 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Philips Azurion systems not configured with an optional auxiliary pan handle. Includes the below product descriptions and corresponding model numbers. 1. Azurion 3 M12; Model Numbers: 722063, 722229. 2. Azurion 3 M15; Model Numbers: 722222, 722064, 722280, 722230. 3. Azurion 5 M12; Model Numbers: 722227, 722231. 4. Azurion 5 M20; Model Numbers: 722228, 722232, 722281. 5. Azurion 7 B12; Model Numbers: 722067, 722225, 722235. 6. Azurion 7 B20; Model Numbers: 722068, 722226, 722236. 7. Azurion 7 M12; Model Numbers: 722078, 722223, 722233. 8. Azurion 7 M20; Model Numbers: 722079, 722224, 722234, 722282.

Why was it recalled?

Potential safety issue with Philips Azurion systems, where longitudinal and transverse table movements may be impacted due to mechanical wear of the Float Tabletop control on the control module.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. Model Number [UDI]: 722063 [(01)00884838085275(21)77, (01)00884838085275(21)47, (01)00884838085275(21)146, (01)00884838085275(21)102, (01)00884838085275(21)43, (01)00884838085275(21)2, (01)00884838085275(21)99, (01)00884838085275(21)28, (01)00884838085275(21)144, (01)00884838085275(21)62, (01)00884838085275(21)119, (01)00884838085275(21)61, (01)00884838085275(21)29, (01)00884838085275(21)40, (01)00884838085275(21)70, (01)00884838085275(21)120, (01)00884838085275(21)121, (01)00884838085275(21)110, (01)00884838085275(21)53, (01)00884838085275(21)103, (01)00884838085275(21)97, (01)00884838085275(21)101, (01)00884838085275(21)143, (01)00884838085275(21)116, (01)00884838085275(21)93, (01)00884838085275(21)141, (01)00884838085275(21)115, (01)00884838085275(21)137, (01)00884838085275(21)150, (01)00884838085275(21)128, (01)00884838085275(21)4, (01)00884838085275(21)50, (01)00884838085275(21)129, (01)00884838085275(21)78, (01)00884838085275(21)76, (01)00884838085275(21)69, (01)00884838085275(21)59, (01)00884838085275(21)92, (01)00884838085275(21)139, (01)00884838085275(21)130, (01)00884838085275(21)154, (01)00884838085275(21)75, (01)00884838085275(21)24, (01)00884838085275(21)94, (01)00884838085275(21)100, (01)00884838085275(21)26, (01)00884838085275(21)106, (01)00884838085275(21)37, (01)00884838085275(21)36, (01)00884838085275(21)27, (01)00884838085275(21)15, (01)00884838085275(21)135, (01)00884838085275(21)123, (01)00884838085275(21)87, (01)00884838085275(21)85, (01)0088483808527

Distribution

Nationwide distribution. International distribution to Albania, Algeria, Argentina, Armenia, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Bolivia, Bosnia and Herzegovina, Brazil, Canada, Chile, C¿te D'Ivoire, El Salvador, Estonia, Greece, Honduras, Hong Kong, Iceland, Indonesia, Iraq, Japan, Kyrgyzstan, Malaysia, Malta, Moldova, Nepal, New Zealand, Philippines, Qatar, Singapore, South Africa, Sri Lanka, Syrian Arab Republic, Tanzania, Turkmenistan, Vietnam, Yemen

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2225-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.