FDA Medical devices recall ·

ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model… recall

Recall number
Z-2206-2026
Date
(initiated March 9, 2026)
Company / brand
Siemens Medical Solutions USA, Inc
40 Liberty Blvd, Malvern, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Initiated by
FDA Mandated

Read the official FDA notice ↗

Product

ARTIS Icono biplane, floor and ceiling.: ARTIS icono biplane Model 11327600. ARTIS icono ceiling Model 11328100. ARTIS icono floor Model 11327700. interventional fluoroscopic x-ray system

Why was it recalled?

During patient examinations, the system may sporadically display a 0 x-ray dose administered to the patient. This behavior has only been observed during fluoroscopy and/or single-frame acquisitions. Despite the reported x-ray dose showing 0, an actual x-ray dose is applied to the patient.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

ARTIS icono biplane UDI 4056869063317 ARTIS icono ceiling UDI 4056869295923 ARTIS icono floor UDI 4056869149325

Distribution

US and Worldwide

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2206-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.