FDA Medical devices recall ·
LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers recall
- Recall number
- Z-2188-2025
- Date
- (initiated May 29, 2025)
- Company / brand
- Microvention Inc.
35 Enterprise, Aliso Viejo, CA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1451
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
LIFEPEARL Drug Elutable Microspheres: 200 +/- 50 micrometers, REF: 8LP2S200; 400 +/- 50 micrometers, REF: 8LP2S400
Why was it recalled?
Drug Elutable Microspheres have a smaller actual average diameter that is not within specification, which may lead to inability to reach the desired treatment location (non-targeted embolization), additional procedure/treatment required, blockage other than target vessel, increased procedure time, incomplete embolization, and/or inability to treat the patient.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF/GTIN/Lot: 8LP2S20/(01)00812636020105/0000177644, 0000180340, 0000223990, 0000235256, 0000254143, 0000254142, 0000272315, 0000272316, 0000281719, 0000294361, 0000295741, 0000296446, 0000751662, 0000751661, 0001026153; 8LP2S400/(01)00812636020112/0000199869, 0000225205, 0000708506, 0000863001, 0001026156
Distribution
OUS: Italy, Spain, Portugal, Poland, Belgium, Germany, Denmark, France, Kazakhstan, Slovakia, Norway, Greece, Serbia, Turkey, Czech Republic, United Kingdom, Netherlands, Russian Federation, Kuwait, Republic of North Macedonia, Switzerland, Saudi Arabia, Georgia, Lithuania, Romania, Croatia, Chile, Hong Kong , India, Mexico, Taiwan, Vietnam
States: Georgia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2188-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.