FDA Medical devices recall ·
BD Alaris Pump Infusion Sets recall
This is an FDA Class I recall — the most serious type.
- Recall number
- Z-2180-2025
- Date
- (initiated July 8, 2025)
- Company / brand
- CareFusion 303, Inc.
10020 Pacific Mesa Blvd, San Diego, CA, United States - Category
- Medical devices › Class I
- Classification
- Class I
- Status
- Ongoing
- Units affected
- 33,162,680 units total
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
BD Alaris Pump Infusion Sets REF: 10013186, BD Alaris Pump Infusion Set, Back Check Valve, 3 SmartSite Y-sites LVP 60D DEHP FREE 3SS CV; REF: 2452-0007, BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 10D 3SS 2CV; REF: 2204-0007, BD Alaris Pump Infusion SetAS LVP 20D; REF: 24001-0007, BD Alaris Pump Infusion Set, Back Check Valve, SmartSite Y-siteAS LVP 20D SS CV; REF: 10015012, BD Alaris Pump Infusion Burette Set, 0.2 Micron Filter, Smallbore Tubing, SmartSite Port (Burette), SmartSite Y-siteAS LVP BUR 20D PE-Lined SMBORE SS 0.2M; REF: 2432-0007, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D 3SS 0.2M CV; REF: 2434-0007, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, SmartSite Y-siteAS LVP 20D SS 0.2M CV; REF: 2202-0007, BD Alaris Pump Infusion Set, 1.2 Micron FilterAS LVP 20D 1.2M; REF: 11426965, BD Alaris Pump Infusion Set, Back Check Valve, 5 SmartSite Y-sitesAS LVP 20D CV; REF: 10561554, BD Alaris Pump Infusion Set, 15 Micron Filter, Back Check Valve, 3 SmartSite Y-sitesAS LVP 20D DEHP FREE 3SS 15M CV; REF: 10863358, BD Alaris Pump Infusion Set, 15 Micron FilterAS LVP 20D DEHP FREE 15M; REF: 11171447, BD Alaris Pump Infusion Set, 2 Back Check Valves, 3 SmartSite Y-sitesAS LVP 20D 3SS 2CV; REF: 11532269, BD Alaris Pump Infusion Set, 0.2 Micron Filter, Back Check Valve, PE Lined Tubing, 2 SmartSite Y-sitesAS LVP 20D PE-LINED 2SS 0.2M CV; REF: 10010454, BD Alaris Pump Infusion Set, 0.2 Micron Filter, PE Lined Tubing, SmartSite Y-siteAS LVP 20D PE-LINED TUBING SS 0.2M; REF: 2465-0007, BD Alaris Pump Infusion Set, SmartSite Bag Access Non-Vented, 0.2 Micron Filter, Low Sorbing Tubing (PE Lined), SmartSite Y-siteAS LVP SS BAG 20D PE-Lined SS 0.2M; REF: 10010453, ASBD Alaris Pump Infusion Set, 1.2 Micron Filter, PE Lined Tubing, SmartSite Y-site LVP 20D PE-LINED TUBING SS 1.2M REF: 11419365, BD Alaris" Pump Infusion Set, Back Check Valve, Manifold, 3-Way Stopcock, 5 SmartSite" Y-sites REF: 10813621, BD Alaris" Pump Infusion Set, Back Check Valve, 3 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 2423-0007, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 4 SmartSite" Y-sites REF: 10015896, BD Alaris" Pump Infusion Set, Back Check Valve, 2 Ganged 4-Way Stopcocks, 3 SmartSite" Y-sites REF: 24010-0007T, BD Alaris" Pump Infusion Set, Bonded Texium", Closed Male Luer with Priming Cap, Back Check Valve, 3 SmartSite" Y-sites REF: 10010483, BD Alaris" Pump Infusion Set, Vented Syringe Adapter, Smallbore Tubing REF: 2403-0007, BD Alaris" Pump Infusion Set, Half Set, SmartSite" Y-site REF: 24301-0007T, BD Alaris" Pump Infusion Set, SmartSite" Bag Access Non-Vented, 0.2 Micron Filter, PE Lined Tubing, Bonded Texium" Closed Male Luer with Priming Cap, SmartSite" Y-site
Why was it recalled?
Infusion pump module used with compatible pump infusion sets may perform outside the established performance ranges for flow rate and bolus accuracy, downstream and upstream occlusion time to alarm, and post-occlusion bolos volume.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF: 10013186/ UDI: 07613203021173; REF: 2452-0007/ UDI: 10885403219870; REF: 2204-0007/ UDI: 10885403199363; REF: 24001-0007/ UDI: 10885403238666; REF: 10015012/ UDI: 10885403233951; REF: 2432-0007/ UDI: 10885403232329; REF: 2434-0007/ UDI: 07613203019682; REF: 2202-0007/ UDI: 10885403274039; REF: 11426965/ UDI: 10885403232558; REF: 10561554/ UDI: 10885403232565; REF: 10863358/ UDI: 10885403232466; REF: 11171447/ UDI: 07613203021234; REF: 11532269/ UDI: 10885403232343; REF: 10010454/ UDI: 07613203015806; REF: 2465-0007/ UDI: 10885403221941; REF: 10010453/ UDI: 07613203021135 REF: 11419365/ UDI: 10885403234736 REF: 10813621/UDI: 10885403234514 REF: 2423-0007/UDI: 07613203020985 REF: 10015896/UDI: 10885403234064 REF: 24010-0007T/UDI: 10885403271021 REF: 10010483/UDI: 10885403232602 REF: 2403-0007/UDI: 10885403232473 REF: 24301-0007T/UDI: 10885403223198
Distribution
US Nationwide Distribution to states of: AK AL AR AS AZ CA CO CT DC DE FL GA GU HI IA ID IL IN KS KY LA MA MD ME MI MN MO MP MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV WY; and OUS (foreign) to countries of: Japan, Italy, Germany, Turkey
States: Alaska Arkansas Arizona California Connecticut District of Columbia Florida Georgia Iowa Illinois Kansas Kentucky Maryland Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming
What does Class I mean?
Class I is the most serious FDA recall class: there is a reasonable probability that using or being exposed to the product will cause serious adverse health consequences or death.
Source: FDA (medical devices), recall Z-2180-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.