FDA Medical devices recall ·

Philips Azurion recall

Recall number
Z-2170-2026
Date
(initiated March 31, 2026)
Company / brand
Philips Medical Systems Nederland B.V.
Veenpluis 6, Best, Netherlands
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1,718 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Philips Azurion. System Code Description (Model Number): Azurion 3 M12 (722221, 722229), Azurion 3 M15 (722222, 722230), Azurion 5 (722281), Azurion 5 M12 (722227, 722231), Azurion 5 M20 (722228, 722232, 722281), Azurion 7 (722282), Azurion 7 B12 (722225, 722235), Azurion 7 B20 (722068, 722226, 722236), Azurion 7 M12 (722078, 722223, 722233), Azurion 7 M20 (722079, 722224, 722234, 722282).

Why was it recalled?

Potential for the bolts and plastic parts of the Cable Hose Carriers that hold the cable hose to the monitor ceiling suspension may loosen and/or break over time due to the forces applied when the monitor is moved or rotated.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

1. System Code Description: Azurion 3 M12. Model Number [UDI]: 722221 [(01)00884838099203(21)181, (01)00884838099203(21)178, (01)00884838099203(21)93]; 722229 [(01)00884838116726(21)22, (01)00884838116726(21)9, (01)00884838116726(21)16, (01)00884838116726(21)15, (01)00884838116726(21)5, (01)00884838116726(21)3, (01)00884838116726(21)4, (01)00884838116726(21)6, (01)00884838116726(21)2, (01)00884838116726(21)8]. 2. System Code Description: Azurion 3 M15. Model Number [UDI]: 722222 [(01)00884838099210(21)358, (01)00884838099210(21)328, (01)00884838099210(21)389, (01)00884838099210(21)378, (01)00884838099210(21)243, (01)00884838099210(21)392, (01)00884838099210(21)396, (01)00884838099210(21)13, (01)00884838099210(21)379, (01)00884838099210(21)385, (01)00884838099210(21)376, (01)00884838099210(21)366, (01)00884838099210(21)365, (01)00884838099210(21)398, (01)00884838099210(21)274]; 722230 [(01)00884838116733(21)40, (01)00884838116733(21)2, (01)00884838116733(21)41, (01)00884838116733(21)15, (01)00884838116733(21)28, (01)00884838116733(21)38, (01)00884838116733(21)9, (01)00884838116733(21)27, (01)00884838116733(21)13]. 3. System Code Description: Azurion 5. Model Number [UDI]: 722281 [(01)00884838110564(21)401, (01)00884838110564(21)427, (01)00884838110564(21)382, (01)00884838110564(21)425]. 4. System Code Description: Azurion 5 M12. Model Number [UDI]: 722227 [(01)00884838099227(21)265, (01)00884838099227(21)360, (01)00884838099227(21)320, (01)00884838099227(21)338, (01)0088483809

Distribution

Worldwide distribution - US Nationwide and the country of Canada.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2170-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.