FDA Medical devices recall ·
Covidien Nellcor Bedside SpO2 Patient Monitoring System recall
- Recall number
- Z-2151-2025
- Date
- (initiated June 23, 2025)
- Company / brand
- Covidien, LP
6135 Gunbarrel Ave, Boulder, CO, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 264,790
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Covidien Nellcor Bedside SpO2 Patient Monitoring System: REF: PM100N, 10005941, DLPM100N, DL10005941, DSPM100N, 10005941-SG, DS10005941, 10005941J, 10005941JP, 10005941JPN
Why was it recalled?
Bedside SpO2 Patient Monitoring System alarms not heard/recognized, resulting in: treatment delay, lack of low oxygen saturation response, respiratory failure, arrhythmia. Highlighting operator's manual: 1)For patient Safety don't silence/decrease alarm volume, 2)Sensor off alarm (no perfusion detected) is low priority alarm, 3)Set monitors to Homecare Mode outside of hospital/professional setting
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
REF/UI-DI/Bundle Code/IFU Part Number: PM100N/10884521196728/NELL-HOMECARE, BNPULSOKSYMETR, PM100N, PM100NKIT, PM100NKITCHDE, PM100NKITCHFR, PM100NKITDA, PM100NKITDE-2, PM100NKITES, PM100NKITFI, PM100NKITFR, PM100NKIT-FR, PM100NKITIT, PM100NKIT-NL, PM100NKITNO, PM100NKITPT, PM100NKITSV, PM100NKITUK, PM100N-10, PM100N-2XMAXN, PM100N-2XMAXN-CC, PM100N-3DYS-CC, PM100N-DYS, PM100N-HC, PM100N-MAXN, PM100N-MAXN-CC, PM100N-OXIAN-CC/PT00156589 - AR - Arabic, PT00156609 - AR - Arabic, PT00156588 - BG - Bulgarian, PT00156610 - BG - Bulgarian, PT00156586 - CS - Czech, PT00156612 - CS - Czech, PT00156585 - DA - Danish, PT00156613 - DA - Danish, PT00156578 - DE - German, PT00156625 - DE - German, PT00156577 - EL - Greek, PT00156626 - EL - Greek, PT00156509 - EN - English, PT00156328 - EN - English, PT00156563 - ES - Spanish, PT00156639 - ES - Spanish, PT00156583 - ET - Estonian, PT00156615 - ET - Estonian, PT00156582 - FI - Finnish, PT00156616 - FI - Finnish, PT00156581 - FR - French, PT00156617 - FR - French, PT00156587 - HR - Croatian , PT00156611 - HR - Croatian , PT00156576 - HU - Hungarian, PT00156627 - HU - Hungarian, PT00156575 - IT - Italian, PT00156628 - IT - Italian, PT00156580 - KA - Georgian, PT00156618 - KA - Georgian, PT00156591 - KO - Korean, PT00156642 - KO - Korean, PT00156573 - LT - Lithuanian, PT00156630 - LT - Lithuanian, PT00156574 - LV - Latvian, PT00156629 - LV - Latvian, PT00156584 - NL - Dutch, PT00156614 - NL - Dutch, PT00156572 - NO - Norwegian, PT00156631 - NO
Distribution
Worldwide - US Nationwide distribution including in the states of MS, TX, FL, SC, VA, AZ, PA, MN, OH, CA, MI, CO, KS, IN, IL, TN, MO, IA, OK, WI, AK, NY, MD, AL, WV, GA, NC, AR, WA, SD, KY, DE, NJ, UT, NV, MA, MT, GU, NE, LA, HI, OR, RI, NH, ND, NM, ME, CT, DC, WY and the countries of Algeria, Andorra, Argentina, Australia, Austria, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Bhutan, Bolivia, Bosnia And Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Cambodia, Canada, Canary Islands, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Denmark, Egypt, Estonia, Finland, France, French Guiana, Germany, Ghana, Greece, Guadeloupe, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Islamic Republic Of, Ireland, Isle of Man, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Lebanon, Libya, Lithuania, Macedonia, Malaysia, Maldives, Malta, Mauritius, Mexico, Montenegro, Morocco, Myanmar, Namibia, Netherlands, New Zealand, North Macedonia, Norway, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Qatar, Reunion, Romania, Russia, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sri Lanka, Sweden, Switzerland, Taiwan, Thailand, Tunisia, Turkey, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Vietnam, Zimbabwe.
States: Alaska Alabama Arkansas Arizona California Colorado Connecticut District of Columbia Delaware Florida Georgia Hawaii Iowa Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin West Virginia Wyoming
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2151-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.