FDA Medical devices recall ·

Discovery MR750w 3.0T recall

Recall number
Z-2132-2025
Date
(initiated May 23, 2025)
Company / brand
GE Medical Systems, LLC
3200 N Grandview Blvd, Waukesha, WI, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
N/A
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Discovery MR750w 3.0T

Why was it recalled?

GE HealthCare has become aware that gradient coils for certain MR systems (see affected product list in this letter), under specific conditions, can produce elevated acoustic noise during scanning. As a result, acoustic levels can exceed 99dB, the limit established by International Electrotechnical Commission (IEC 60601-2-33) when using hearing protection with a Noise Reduction Rating (NRR) of 29dB, as currently required in the Operator Manual.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

GTIN: 00840682103817 00195278229519

Distribution

Worldwide distribution - US Nationwide and the country of China.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2132-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.