FDA Medical devices recall ·
VITROS Chemistry Products Product Name: VITROS Chemistry Products… recall
- Recall number
- Z-2120-2025
- Date
- (initiated June 10, 2025)
- Company / brand
- Ortho-Clinical Diagnostics, Inc.
100 Indigo Creek Dr, Rochester, NY, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 37429 units (US)/11,703 units (OUS)
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: VITROS Chemistry Products Product Name: VITROS Chemistry Products Performance Verifier I Model/Catalog Number: 806 7324 Software Version: Not Applicable Product Description: VITROS Chemistry Products Performance Verifier is an assayed control used to monitor performance on VITROS 250/350/5,1 FS/4600/XT 3400 Chemistry Systems and the VITROS 5600/XT 7600 Integrated Systems. VITROS Performance Verifier is prepared from processed human serum to which enzymes, electrolytes, stabilizers,mpreservatives and other organic analytes have been added. Component: No, not a component
Why was it recalled?
Quality control (QC) drift when certain lots of Performance Verifiers (PVs) are used with certain lots of cholinesterase enzyme (CHE) slides. This drift occurs when the PVs are used 2-7 days after reconstitution; however, the label stability claims for the PV fluids is 7 days following reconstitution. This may lead to inaccurate QC results. The use of the PVs when the QC results are outside of the customer s established ranges could result in delayed results for patients due to the need to repeat QC testing or reconstitute new PV fluids and repeat testing. It could also result in erroneous patient results (small positive bias).
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Lot Code: Product Code: 806 7324 (UDI: 10758750004317) Lots currently within expiry: V1665 - Exp.8/14/2025 B2245 - Exp 3/25/2026 Y2110 - Exp. 3/25/2026 E2828 - Exp. 2/3/2027
Distribution
Worldwide Distribution: US (nationwide) to states of: AZ, CA, FL, GA, IL, MD, MI, MO, MS, NY, TN, TX, VA; and OUS (International) to countries of: Australia, Brazil, Canada, Chile, China, Colombia, Denmark, France, Germany, India, Italy, Japan, Mexico, Netherlands, Norway, Portugal, Russia, Singapore, Spain, Sweden, and United Kingdom.
States: Arizona California Florida Georgia Illinois Maryland Michigan Missouri Mississippi New York Tennessee Texas Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2120-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.