FDA Medical devices recall ·

NIO+ Product Name: NIO Intraosseous Device Adult Model recall

Recall number
Z-2119-2025
Date
(initiated June 5, 2025)
Company / brand
Waismed Ltd.
11, Galgalei Ha-Plada, Herzliya, N/A, Israel
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
516 units (US)
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: NIO+ Product Name: NIO Intraosseous Device Adult Model/Catalog Number: 105000908 Software Version: N/A Product Description: NIO devices are a spring-based, automatic intraosseous access device supplied Sterile and single use. Indicated for Intraosseous access to the Proximal humerus and Proximal Tibia in adult patients, in emergent situations. Component: N/A

Why was it recalled?

Due to manufacturing error, Intraosseous device built-in stabilizer may not be released properly after deployment and insertion.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: UDI: 07290008325073 Lot number: 2440044

Distribution

Worldwide distribution: US (nationwide) to state of: NC and OUS (international) countries of: Slovenia, Australia, Brazil, Estonia, South Africa, U. Arab Emirate, Greece, Panama, Singapore, Italy, Germany, Portugal, Spain, Ukraine, Lithuania, Chile, Poland, Germany, New Zealand, Montenegro, Switzerland, Argentina, Poland, Austria, Sweden, United Kingdom, Italy, France, Romania, Netherlands, Croatia, Lithuania, New Zealand, Japan, Cyprus, and Israel.

States: North Carolina

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2119-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.