FDA Medical devices recall ·

SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve recall

Recall number
Z-2116-2025
Date
(initiated June 3, 2025)
Company / brand
ICU Medical, Inc.
951 Calle Amanecer, San Clemente, CA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1,445,600
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

SwabCap Disinfecting Cap for Needlefree Connectors: SwabSleeve, REF: SCXT3-5-2000; SwabCap, 200-ct. Box, REF: SCXT3-2000

Why was it recalled?

Due to a manufacturing issue, disinfecting cap for needle-free connectors may have an incomplete seal between the foil lid and plastic container, which may result in isopropyl alcohol evaporation from the sponge, which may result in inadequate disinfection.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

REF/UDI-DI/Lot(Expiration): SCXT3-5-2000/10887709097107/14258739(1-Dec-26), 14258743(1-Feb-26), 14258744(1-Dec-26), 14264200(1-Dec-26), 14272241(1-Dec-26), 14272248(1-Dec-26), 14287456(1-Dec-26), 14287459(1-Dec-26), 14287480(1-Dec-26), 14287482(1-Dec-26), 14287485(1-Dec-26), 14294832(1-Dec-26), 14272254(1-Dec-26), 14310313(1-Dec-26), 14310425(1-Dec-26), 14323354(1-Dec-26), 14287475(1-Dec-26), 14294840(1-Dec-26), 14321645(1-Jan-27), 14321652(1-Jan-27), 14331169(1-Jan-27), 14332773(1-Jan-27), 14341732(1-Oct-26), 14294825(1-Jan-27), 14303209(1-Jan-27), 14310308(1-Jan-27), 14355404(1-Jan-27), 14355409(1-Jan-27) SCXT3-2000/20887709079162/14258214(1-Jan-27), 14258206(1-Jan-27), 14258208(1-Feb-27).

Distribution

Worldwide - US Nationwide distribution including in the states of TX, SD, ID, LA, IN, CO, OK, MO, MN, OR, AZ, NC, TN, VA, CA, IL, FL, MA, OH, MD, PA, NJ, NY, WA, GA, UT, MI, MS, AR, AL, NV, KS, IA, WV, HI, KY, NM and the country of CA.

States: Alabama Arkansas Arizona California Colorado Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi North Carolina New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania South Dakota Tennessee Texas Utah Virginia Washington West Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2116-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.