FDA Medical devices recall ·

Laser light show projectors manufactured by BLS, Model CLUB 1000/2000/4000 recall

Recall number
Z-2112-2025
Date
(initiated June 30, 2025)
Company / brand
Beyond Laser Systems, LLC
4211 Todd Ln Ste A, Austin, TX, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
12
Initiated by
FDA Mandated

Read the official FDA notice ↗

Product

Laser light show projectors manufactured by BLS, Model CLUB 1000/2000/4000.

Why was it recalled?

The remote interlock connector, which is required by 21 CFR 1040.10(f)(3), in the remote stop box was not installed the connector or was mechanically installed but not wired. The labels placed on the projectors do not meet requirements specified in 21 CFR 1010.4 (Variances), 21 CFR 1010.2 (Certification), 21 CFR 1010.3 (Identification) and 21 CFR 1040.10 (g)(Labelling requirements). The identification label does not include the date of manufacture as required by 21 CFR 1010.3. The corrective action plan covers three families of laser light show of projectors: DIODE. CLUB and PURE with three different hardware configurations such as the number of the number of RJ45 ports, radiant power of emissions and network configurations. No specific user manuals were provided to each model family as required by 21 CFR 1040.10(h)(1) (Informational requirements (1) User information). The Quality Control test did not validate the manual reset function to enable resumption of laser radiation emission from Class IV laser projectors after an interruption caused using a remote interlock. No supplemental product reports, as required according to 21 CFR 1002.11.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

N/A

Distribution

US Nationwide Distribution

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2112-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.