FDA Medical devices recall ·
Medline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A recall
- Recall number
- Z-2102-2026
- Date
- (initiated January 7, 2026)
- Company / brand
- Medline Industries, LP
3 Lakes Dr, Northfield, IL, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1516 units
- Hazard tags
- Sterility
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Medline Convenience Kits: 1) CVC PACK, Model Number: CVI4500A; 2) TOTE KAMAL OSC HAND REPAIR, Model Number: DYKM1730A; 3) TOTE CHANGE OSC HAND REPAIR, Model Number: DYKM1751A; 4) TOTE SB NGUYEN CAPSULOTOMY, Model Number: DYKM1756A; 5) TOTE MOMENI SB CAPSULOTOMY REC, Model Number: DYKM1765A; 6) TOTE SAFRAN OSC HIP ARTHRSCPY, Model Number: DYKM1767B; 7) TOTE SEN OSC HAND REPAIR, Model Number: DYKM1776; 8) TOTE CURTIN OSC HAND REPAIR, Model Number: DYKM1781A; 9) TOTE FOX OSC HAND REPAIR, Model Number: DYKM1783A; 10) TOTE YAO OSC HAND REPAIR, Model Number: DYKM1796A; 11) TOTE ABRAMS KNEE ARTHROSCOPY, Model Number: DYKM1801B; 12) TOTE YAO OSC WRIST ARTHROSCOPY, Model Number: DYKM1807; 13) TOTE CARROLL BLOCK BLOOD PATCH, Model Number: DYKM1864; 14) TOTE SB STONE JWIRE LUMPECTOMY/SEN NODE, Model Number: DYKM1925; 15) TOTE SB NGUYEN LIPOSUCTION, Model Number: DYKM1927A; 16) OSC SHERMAN KNEE ARTHRO BASIC, Model Number: DYKM1977B; 17) CHEST TUBE SUPPLIES, Model Number: DYKM2131
Why was it recalled?
Medline has identified issues related to calibration of the equipment used to sterilize and package the devices. All products were exposed to the validated sterilization and packaging cycles; however, the identified calibration issues have the potential to impact the sterility assurance level (SAL) of the Recalled Products.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
1) CVI4500A, UDI-DI: 10653160327646(each), 653160327649(case), Lot Number: 2020121550; 2) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024030690; 3) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024052290; 4) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2024060590; 5) DYKM1730A, UDI-DI: 10193489111019(each), 20193489111016(case), Lot Number: 2025012890; 6) DYKM1751A, UDI-DI: 10193489111002(each), 20193489111009(case), Lot Number: 2023121190; 7) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023060290; 8) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023070590; 9) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023111590; 10) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2023120590; 11) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024022190; 12) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024080190; 13) DYKM1756A, UDI-DI: 10193489190250(each), 20193489190257(case), Lot Number: 2024101490; 14) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2023050490; 15) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2023071290; 16) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190233(case), Lot Number: 2024012490; 17) DYKM1765A, UDI-DI: 10193489190236(each), 20193489190
Distribution
Worldwide - US Nationwide distribution.
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-2102-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.