FDA Medical devices recall ·
The HemoCue Glucose 201 Microcuvettes are designed for use with the… recall
- Recall number
- Z-2070-2025
- Date
- (initiated May 12, 2025)
- Company / brand
- HemoCue AB
Kuvettgatan 1, Angelholm, Sweden - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 4 Boxes
- Hazard tags
- Potency
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
The HemoCue Glucose 201 Microcuvettes are designed for use with the HemoCue Glucose 201 (or DM) Analyzer with plasma conversion that multiplies the measured whole blood glucose value by a factor of 1.112 and displays a plasma equivalent glucose result. HemoCue Glucose 201 Microcuvettes are available in individual packages or in vials. The box contains 4 vials with 25 microcuvettes in each vial.
Why was it recalled?
Glucose microcuvettes experienced transit time outside of limits, so stability through the whole microcuvette lifetime can no longer be guaranteed. The affected products may not have been stored at the recommended labeled storage conditions which may have impacted the safety, quality, identity, potency, and purity of the product. This failure could increase the probability of an incorrect test result or a delayed test result.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI: 07311091107060, Lot: 2501708 (Only 4 boxes within specific McKesson Shipment)
Distribution
US Nationwide distribution in the state of AR.
States: Arkansas
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.