FDA Medical devices recall ·

CDI OneView System BPM Probe: Cat recall

Recall number
Z-2018-2025
Date
(initiated May 28, 2025)
Company / brand
Terumo Cardiovascular Systems Corporation
6200 Jackson Rd, Ann Arbor, MI, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
28 US; 7 OUS
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

CDI OneView System BPM Probe: Cat. No. CDI753 - CDI OneView BPM Probe, Cat. No. CDI753TK - CDI OneView BPM Probe Trial Kit

Why was it recalled?

Due to an error in assembly of the thermistor within some units, the affected probes will experience inaccuracy in measured shunt sensor temperature. Additionally, following changes in patient temperature without an in vivo recalibration, affected probes may experience inaccuracy in other measured parameters, including pCO2, pO2, pH, Potassium and calculated parameters, including BE, Bicarbonate, Arterial Oxygen Saturation, Delivered Oxygen, Area Under of the Curve of Delivered Oxygen, Oxygen Consumption, and Oxygen Extraction Ratio, which utilize temperature as an input. Inaccurate data may lead to inappropriate user response.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI 00886799001889 Lot Number/Serial Number Range: B0050231-B0050240 B0050221-B0050230 B0050211-B0050219 B0050201-B0050209 B0050191-B0050200

Distribution

US Nationwide distribution in the states of California, Delaware, Hawaii, Iowa, Indiana, Michigan, North Carolina, New York, Ohio, and Washington. The country of New Zealand.

States: California Delaware Hawaii Iowa Indiana Michigan North Carolina New York Ohio Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-2018-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.