FDA Medical devices recall ·

ICU Medical Plum Solos, Item number 400011001 recall

Recall number
Z-1991-2025
Date
(initiated April 22, 2025)
Company / brand
ICU Medical, Inc.
600 N Field Dr, Lake Forest, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
214 units
Hazard tags
Software
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

ICU Medical Plum Solos, Item number 400011001; infusion pump

Why was it recalled?

Internal testing found that Plum Duo pumps with software version 1.1.1 have a workflow that may result in bypassing the Maximum Dose Limit alert. The Maximum Dose Limit is set in the LifeShield Drug Library Manager (DLM) within the LifeShield Medication Management Safety Software and specifies the highest dose at which the pump can be programmed for weight or BSA (body surface area) based medication rulesets. The limit can be configured for Dose, Loading Dose, and Bolus Dose.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI/DI N/A, Serial/Lot Numbers: 60000001, 60000002, 60000003, 60000005, 60000006, 60000007, 60000008, 60000009, 60000010, 60000011, 60000014, 60000015, 60000016, 60000017, 60000019, 60000020, 60000021, 60000023, 60000024, 60000025, 60000026, 60000027, 60000028, 60000030, 60000032, 60000033, 60000035, 60000038, 60000039, 60000040, 60000041, 60000042, 60000044, 60000045, 60000046, 60000047, 60000048, 60000049, 60000050, 60000051, 60000052, 60000053, 60000054, 60000056, 60000057, 60000058, 60000059, 60000060, 60000061, 60000062, 60000063, 60000064, 60000065, 60000066, 60000067, 60000068, 60000069, 60000070, 60000071, 60000072, 60000073, 60000074, 60000075, 60000076, 60000077, 60000078, 60000079, 60000080, 60000081, 60000082, 60000083, 60000084, 60000085, 60000087, 60000088, 60000089, 60000090, 60000092, 60000093, 60000094, 60000095, 60000096, 60000097, 60000098, 60000099, 60000100, 60000101, 60000102, 60000104, 60000105, 60000107, 60000108, 60000109, 60000110, 60000112, 60000113, 60000114, 60000120, 60000130, 60000133, 60000134, 60000136, 60000139, 60000141, 60000143, 60000147, 60000150, 60000154, 60000159, 60000160, 60000162, 60000165, 60000167, 60000170, 60000171, 60000172, 60000173, 60000174, 60000176, 60000177, 60000178, 60000179, 60000180, 60000183, 60000187, 60000189, 60000190, 60000191, 60000192, 60000198, 60000199, 60000201, 60000203, 60000204, 60000205, 60000206, 60000207, 60000208, 60000209, 60000210, 60000212, 60000213, 60000214, 60000215, 60000216, 60000217, 60000223

Distribution

Worldwide - US Nationwide distribution in the states of FL, KY, VA, MS, MA, PA, NY, CA, IA and the country of Philippines.

States: California Florida Iowa Kentucky Massachusetts Mississippi New York Pennsylvania Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1991-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.