FDA Medical devices recall ·

UROSKOP Omnia Max recall

Recall number
Z-1982-2025
Date
(initiated June 4, 2025)
Company / brand
Siemens Medical Solutions USA, Inc
40 Liberty Blvd, Malvern, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
25 units
Hazard tags
Fire
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

UROSKOP Omnia Max. Model Number: 10762473

Why was it recalled?

A resistor in the frequency inverter may strongly overheat potentially igniting the plastic housing of the frequency inverter on fire.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model Number: 10762473. UDI Numbers: (01)04056869008981(21)4014, (01)04056869008981(21)4056, (01)04056869008981(21)4049, (01)04056869008981(21)4015, (01)04056869008981(21)4063, (01)04056869008981(21)4022, (01)04056869008981(21)4036, (01)04056869008981(21)4043, (01)04056869008981(21)4024, (01)04056869008981(21)4059, (01)04056869008981(21)4020, (01)04056869008981(21)4032, (01)04056869008981(21)4053, (01)04056869008981(21)4061, (01)04056869008981(21)4060, (01)04056869008981(21)4051, (01)04056869008981(21)4025, (01)04056869008981(21)4037, (01)04056869008981(21)4054, (01)04056869008981(21)4042, (01)04056869008981(21)4033, (01)04056869008981(21)4055, (01)04056869008981(21)4052, (01)04056869008981(21)4047, (01)04056869008981(21)4062. Serial Numbers: 4014, 4056, 4049, 4015, 4063, 4022, 4036, 4043, 4024, 4059, 4020, 4032, 4053, 4061, 4060, 4051, 4025, 4037, 4054, 4042, 4033, 4055, 4052, 4047, 4062.

Distribution

Nationwide distribution.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1982-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.