FDA Medical devices recall ·

Dimension Creatinine Flex reagent cartridge recall

Recall number
Z-1980-2026
Date
(initiated March 18, 2026)
Company / brand
Siemens Healthcare Diagnostics, Inc.
500 Gbc Dr, Newark, DE, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
18,080 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Dimension Creatinine Flex reagent cartridge. Material Number: 10872079.

Why was it recalled?

Siemens Healthcare Diagnostics is recalling certain lots of Dimension Creatinine Flex reagent cartridge (Dimension CRE2) because a potential imprecision in the Dimension CRE2 quality control (QC) and patient sample results using lot numbers GA6307 and BA7005 on the Dimension System. This imprecision may lead to erroneously decreased or increased creatinine patient results

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Material Number: 10872079. Model Number: DF33B. UDI: (01)00630414595009(10)GA6307(17)20261103, (01)00630414595009(10)BA7005(17)20270105. Lot Numbers: GA6307, BA7005.

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Bahamas, Brazil, Cambodia, Canada, Chile, China, Columbia, Costa Rica, Ecuador, Egypt, Germany, Honduras, India, Indonesia, Korea, Kuwait, Lebanon, Malaysia, Mexico, Nicaragua, Pakistan, Paraguay, Philippines, Qatar, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Taiwan, Thailand, U.A.E., Ukraine, Uruguay, Vietnam, Yemen.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1980-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.