FDA Medical devices recall ·

Genius" Review Station Product Name: Genius" Review Station Display (Barco… recall

Recall number
Z-1949-2026
Date
(initiated March 23, 2026)
Company / brand
Hologic, Inc
250 Campus Dr, Marlborough, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1200 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Genius" Review Station Product Name: Genius" Review Station Display (Barco Monitor) Model/Catalog Number: CMP-01669 Software Version: N/A - not software version specific Component: Yes, the Genius" Review Station is a component of the Genius" Digital Diagnostics System with the Genius" Cervical AI Algorithm cleared under DEN210035.

Why was it recalled?

It was identified that certain devices were being operated outside the validated and FDA-cleared display configuration due to user modification of the manufacturer-installed settings. Modifications included disabling Barco Application Appearance Manager (AAM), altering calibration and compliance parameters, and adjusting additional display-related settings. This field correction is being conducted to restore affected system to the validated configuration and reinforce proper use in accordance with the Operator's Manual.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model Number: CMP-01669; UDI-DI: 15420045514201; Serial Number: All;

Distribution

Worldwide distribution - US Nationwide and the countries of Asia Pacific, Europe, and the Middle East.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1949-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.