FDA Medical devices recall ·

Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser… recall

Recall number
Z-1923-2025
Date
(initiated May 6, 2025)
Company / brand
Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
2,675 units (190 US, 2485 OUS)
Hazard tags
Burn
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Powered Laser Surgical Instrument Product Name: Soltive Reusable Laser Fibers Model/Catalog Number: TFL-FBX150R, TFL-FBX200R, TFL-FBX365R, TFL-FBX550R, TFL-FBX940R, TFL-FBX940R-JA (Japan Only), TFL-FBX365R-JA (Japan Only), TFL-FBX200BS-JA (Japan Only), TFL-FBX150R-JA (Japan Only), TFL-FBX200R-JA (Japan Only), TFL-FBX550R-JA (Japan Only); Product Description: SOLTIVE Laser Fibers are delivery devices that transmit laser energy from the laser console to the treatment site through the fiber tip. The laser connector secures the device to the laser.

Why was it recalled?

Potential for thermal incident including an unexpected occurrence of smoke, sparks, burning, or unsteady flames during clinical use. The fracturing of the fiber is the cause of, or contributes to, these thermal incidents.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: All lot numbers for the following models: Model UDI: 5 Pack / 1 Piece: TFL-FBX150R UDI: 00821925043893 / 00821925043909 TFL-FBX200R UDI: 00821925043954 / 00821925043961 TFL-FBX365R UDI: 00821925044012 / 00821925044029 TFL-FBX550R UDI: 00821925044050 / 00821925044067 TFL-FBX940R UDI: 00821925044104 / 00821925044098 TFL-FBX940R-JA (Japan Only); TFL-FBX365R-JA (Japan Only); TFL-FBX200BS-JA (Japan Only); TFL-FBX150R-JA (Japan Only); TFL-FBX200R-JA (Japan Only); TFL-FBX550R-JA (Japan Only);

Distribution

Worldwide distribution - US Nationwide and the countries of AU, CA, CL, DE, HK, JP, SG.

States: California Delaware

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1923-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.