FDA Medical devices recall ·

Sample probe sucks a sample dispensed into a tube or cup and discharges… recall

Recall number
Z-1907-2025
Date
(initiated April 30, 2025)
Company / brand
Beckman Coulter Mishima K.K.
454-32, Sunto-Gun, N/A, Japan
Category
Medical devices › Class II
Classification
Class II
Status
Completed
Units affected
10,176
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 179433670

Why was it recalled?

A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Device: REF/UDI-DI/Serial-Lot: AU480: B11810/14987666540664, B12183/14987666540671, B80091/16954701010015, B96692/14987666542965, B96693/14987666542972, C02654/14987666543283, C02655/14987666543290, C02845/16954701010022, C41919/14987666544174, C41920/14987666544181, N3659700/14987666535257, N3659800/14987666535264, N3659900/14987666535271, N3660000/14987666535288, N3660100/14987666535547, N3660200/14987666535554, N3660300/14987666535554, N3660400/14987666535769, N3660500/14987666536032, N3660600/14987666536049, N3660800/14987666536063, N3912400/14987666537084, N3912500/14987666537091, N3150700, N3151000, N3151300, N3151400, N3151500, N3151600, N3151900/ Serial: From 10155 to 2025021482. Sample Probe: MU993400/04987666522939, Lots: From 178713114 to 179433670. B66751 Kits containing MU993400, Lots: From 178713114 to 179433670

Distribution

Worldwide Distribution: US (nationwide) including states of: CA, PA, VA, TX, NJ, FL, IN, MA, NY, NC, OH, LA, AZ, MO, WA, GA, UT, KY, AL, CO, IL, NE, OK, NV, SC, TN, MD, OR, AR, MS, AK, WI, DE, MT, HI, KS, ME, NH, WV, NM, CT, IA, MI, MN, ID, RI, SD, WY, ND, DC, VT, GUAM, PR and OUS (foreign) to countries of: Viet Nam, Japan, India, Australia, El Salvador, China, United Arab Emirates, Thailand, Russian Federation, Italy, Germany, Korea (Republic of, Malaysia), Taiwan, Province of China, Turkey, Kazakhstan, Israel, Ecuador, Hungary, Egypt, Serbia, Bangladesh, Bosnia and Herzegovina, Costa Rica, Belarus, United Kingdom, Ireland, Botswana, Kenya, Mauritius, Ghana, Zambia, South Africa, Libya, Croatia, France, Switzerland, Czechia, Philippines, Namibia, Canada, Qatar, Azerbaijan, Lesotho, Bulgaria, Zimbabwe, Poland, Norway, Spain, Brazil, Tanzania (United Republic of), Lithuania, Yemen, Uzbekistan, Algeria, Somalia, Sri Lanka, Nepal, Pakistan, Ukraine, Peru, Morocco, Singapore, Slovakia, Mexico, Myanmar, Lebanon, Jordan, Malawi, Greece, Indonesia, Chile, Romania, Georgia, Ethiopia, Sweden, Albania, Eswatini, Nigeria, Cote d'Ivoire, Argentina, Dominican Republic, Iraq, Maldives, Saudi Arabia, Kuwait, Austria, Eritrea, Bolivia (Plurinational State of), Tunisia, Curacao, Mozambique, Hong Kong, Bahrain, Portugal, Panama, New Zealand, Colombia, Uganda, Oman, Belgium, Saint Pierre and Miquelon, Guyana, Macao, Cabo Verde, Sint Maarten (Dutch part), Liberia, Moldova, Republic of, North Macedonia, Congo (Democratic Republic of the), Uruguay, Cyprus, Netherlands, Suriname, Slovenia, Denmark, Finland, Venezuela, Sierra Leone

States: Alaska Alabama Arkansas Arizona California Colorado Connecticut District of Columbia Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Louisiana Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico Rhode Island South Carolina South Dakota Tennessee Texas Utah Virginia Vermont Washington Wisconsin West Virginia Wyoming

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1907-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.