FDA Medical devices recall ·

Philips AneurysmFlow recall

Recall number
Z-1805-2026
Date
(initiated March 19, 2026)
Company / brand
Philips Medical Systems Nederland B.V.
Veenpluis 4-6, Best, N/A, Netherlands
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
1,033 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Philips AneurysmFlow. Model Number: 001015. AneurysmFlow is a software medical device (Interventional Tool) intended to be used in combination with a Philips interventional X-ray system and 3DRA data.

Why was it recalled?

Potential safety issue involving the Mean Aneurysm Flow Amplitude (MAFA) ratio of AneurysmFlow (Interventional Tool). Philips has determined that the MAFA ratio does not provide reliable prognostic information regarding aneurysm occlusion following Flow Diverter Stent treatment, however despite being stated in the IUF, it is being relied on for making clinical decisions.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model Number: 001015. Software Version Number R1.x. UDI Number: (01)00884838099272(21)153, (01)00884838099272(21)67, (01)00884838099272(21)675, (01)00884838085367(21)178, (01)00884838099272(21)112, (01)00884838085367(21)335, (01)00884838116801(21)345, (01)00884838085367(21)147, (01)00884838085367(21)205, (01)00884838099272(21)113, (01)00884838099272(21)121, (01)00884838116801(21)24, (01)00884838085367(21)170, (01)00884838099258(21)1150, (01)00884838116801(21)17, (01)00884838116801(21)25, (01)00884838059061(21)639, (01)00884838085268(21)856, (01)00884838116801(21)180, (01)00884838116801(21)227, (01)00884838099272(21)239, (01)00884838099272(21)105, (01)00884838116801(21)235, (01)00884838099272(21)678, (01)00884838054202(21)373, (01)00884838059115(21)290, (01)00884838085367(21)564, (01)00884838085268(21)716, (01)00884838085367(21)440, (01)00884838116801(21)226, (01)00884838099272(21)197, (01)00884838099272(21)380, (01)00884838059115(21)274, (01)00884838085367(21)247, (01)00884838085367(21)225, (01)00884838099272(21)592, (01)00884838099272(21)593, (01)00884838099272(21)319, (01)00884838099272(21)339, (01)00884838099272(21)182, (01)00884838099258(21)1234, (01)00884838085367(21)64, (01)00884838116801(21)20, (01)00884838099272(21)272, (01)00884838099272(21)680, (01)00884838085367(21)153, (01)00884838099272(21)350, (01)00884838099272(21)280, (01)00884838116801(21)208, (01)00884838116801(21)205, (01)00884838116801(21)36, (01)00884838085367(21)85, (01)00884838085367(21)89, (01)00884838

Distribution

Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Austria, Azerbaijan, Bahrain, Belarus, Belgium, Bnis insted Us Instas instes, Brazil, Canada, Chile, Colombia, Croatia, Czech Republic, Egypt, Estonia, Germany, Greece, Indonesia, Iraq, Italy, Japan, Kuwait, Latvia, Macao, Maldives, Mongolia, Morocco, New Zealand, Nicaragua, Poland, Portugal, Romania, Saudi Arabia, Serbia, Slovenia, South Africa, Sri Lanka, Switzerland, TURKEY, Ukraine, Utd.Arab.Emir., Uzbekistan, Vietnam.

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1805-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.