FDA Medical devices recall ·
Multi-parameter Command Module, Model 91496… recall
- Recall number
- Z-1793-2026
- Date
- (initiated February 27, 2026)
- Company / brand
- Spacelabs Healthcare, Inc.
35301 Se Center St, Snoqualmie, WA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 1790
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Multi-parameter Command Module, Model 91496, optioned with Masimo or Nellcor
Why was it recalled?
Multi-parameter command module has a circuit board issue and when in Auto mode, Cardiac Output (CO) measurement may activate early, resulting in Bad Curve error with no CO measurement or Irregular Curve error with CO measurement not automatically excluded, which may result in inaccurate readings, or the need of additional readings, which may lead to delayed care and/or fluid overload.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Serial Numbers: UDI-DI: 10841522106415, 10841522106422. 1496-223736, 1496-223737, 1496-223738, 1496-224375, 1496-224376, 1496-224400, 1496-224401, 1496-224402, 1496-224403, 1496-224404, 1496-224405, 1496-224406, 1496-223850, 1496-223851, 1496-223852, 1496-224448, 1496-224956, 1496-224957, 1496-224958, 1496-224959, 1496-224960, 1496-224961, 1496-224981, 1496-224258, 1496-223595, 1496-223596, 1496-223597, 1496-223598, 1496-223619, 1496-223620, 1496-223621, 1496-223622, 1496-223623, 1496-223624, 1496-224717, 1496-224718, 1496-223854, 1496-224373, 1496-224374, 1496-223894, 1496-223895, 1496-223896, 1496-223897, 1496-223898, 1496-223899, 1496-223900, 1496-223901, 1496-223902, 1496-223903, 1496-223904, 1496-223905, 1496-223906, 1496-223907, 1496-223918, 1496-225043, 1496-225044, 1496-225045, 1496-225046, 1496-225047, 1496-225048, 1496-225049, 1496-225050, 1496-225051, 1496-225052, 1496-225053, 1496-225054, 1496-225055, 1496-225056, 1496-225057, 1496-225058, 1496-225059, 1496-225060, 1496-225061, 1496-225062, 1496-225063, 1496-225064, 1496-225065, 1496-225066, 1496-225067, 1496-225068, 1496-224949, 1496-223760, 1496-223761, 1496-225154, 1496-224639, 1496-224719, 1496-224720, 1496-224721, 1496-225088, 1496-225089, 1496-224672, 1496-223908, 1496-223909, 1496-223910, 1496-223911, 1496-223912, 1496-223913, 1496-223914, 1496-223915, 1496-223916, 1496-223917, 1496-224637, 1496-224638, 1496-224752, 1496-224744, 1496-224745, 1496-224746, 1496-224747, 1496-224407, 1496-224408, 1496-224409, 1
Distribution
Worldwide - US Nationwide distribution including in the states of NY, NJ, AL, OK, IA, MT, WY, MN, TX, WI, PA, WA, OR, MI, CA, SC, ID, FL, MS, KS, SD, LA, OH, MO, NH, GA, MD, TN, AZ, PR and the countries of TAIWAN, CANADA, KUWAIT, FRANCE, SAUDI ARABIA, POLAND, PAKISTAN, MEXICO, ARGENTINA, PANAMA.
States: Alabama Arizona California Florida Georgia Iowa Idaho Kansas Louisiana Maryland Michigan Minnesota Missouri Mississippi Montana New Hampshire New Jersey New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico South Carolina South Dakota Tennessee Texas Washington Wisconsin Wyoming
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1793-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.