FDA Medical devices recall ·
SIGNA Premier systems recall
- Recall number
- Z-1767-2026
- Date
- (initiated February 27, 2026)
- Company / brand
- GE Healthcare
3001 W Radio Dr, Florence, SC, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 41
- Hazard tags
- Injury risk
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
SIGNA Premier systems
Why was it recalled?
Whole-Body MR scanners used to produce images of the inside of the human body to aid diagnosis of disease may have been manufactured/serviced with ferrous (steel) fittings instead of non-ferrous (brass) fittings on the magnet rear and if system is accessed to perform service while the magnet is ramped, the magnetic field could attract the ferrous fittings potentially resulting in personnel injury.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
UDI-DI/GTIN: 00840682135269, 00195278010797. System ID: 415353UMR2, 415723SHMR19, 413253MR3T, 908725MR, 609652MR3T, M48823, 7168453T, 513584MR5, 513636EASTMR, 513636T1MR, 605333MRPRM, 832522CRMR1, 703858PRE2, 804675VA3T, 082427110162, 082427220143, 082427020147, 082427200242, 082427160461, 082427250139, 082427140382, M4216469, 353745MR01, GON5655515, KRMR56474971, KRMR56203791, KRMR55433661, KRMR56344161, KRMR55774051, KRMR55914661, NOMRI26833321, 15974045, GON1882380, ESMRI27056983, ESMRI26960129, A203MR06, 690120MR01, 11524261, VNMRI26792525, VNMRI26592248, VNMRI26809677
Distribution
Worldwide - US Nationwide distribution in the states of CA, MA, NJ, NY, OH, SD, TX, VA, PR and the countries of Sweden, Switzerland, United Kingdom, Vietnam, China, Norway, Portugal, Spain, France, Germany, Greece, Korea, Republic of
States: California Massachusetts New Jersey New York Ohio Puerto Rico South Dakota Texas Virginia
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1767-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.