FDA Medical devices recall ·
Philips Tempus Pro Patient Monitor recall
- Recall number
- Z-1729-2025
- Date
- (initiated April 15, 2025)
- Company / brand
- Remote Diagnostic Technologies Ltd.
Farnborough Aerospace Centre, Farnborough, N/A, United Kingdom - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 6961
- Hazard tags
- Software
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Philips Tempus Pro Patient Monitor, REF: 00-1004-R, 00-1007-R, 00-1024-R, 00-1026-R, 00-1004, 00-1007
Why was it recalled?
Patient Monitor software update to address issues:1)May freeze at startup with MPM Application error and Attention Restart Required messages,2)Monitor may freeze and reboot at start of 12-lead ECG or after, 3)May reboot after intermittent connection with Smart Mount, 4) 12-lead ECG measurements with Louvain algorithm and software before vx.36 could produce inaccurate rhythm/morphology statements.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Software versions prior to vx.40. REF/UDI-DI: 00-1004-R/05060472440020, 05060472442901; 00-1007-R/05060472440013, 05060472442918; 00-1024-R/05060472441027, 05060472442925; 00-1026-R/05060472441058, 05060472442932; 00-1004/05060472440624; 00-1007/05060472440655
Distribution
Worldwide - US Nationwide distribution including in the states of AZ, VA, FL, CA, TX, NM, IA, NJ, TN, OH, KS, IN, NC, IL, WY, WI, NV, NY, NE, PA, AL, , MA, GA, MO, CO, OK, KY, OR, MS, MI, HI, WA, UT, WV, ID, ND, NH, MN, MT, ME, MD, DE, DC and the countries of AU, GB, IT, SA, DE, AE, BO, FR, ES, BE, IL, SG, CA, IE, TH, NL, MY, AT, DK, RO, CH, IS, PT, NO, HK, PL, KY, OM, CZ, SI, SL, ZA, AF, SE, FI.
States: Alabama Arizona California Colorado District of Columbia Delaware Florida Georgia Hawaii Iowa Idaho Illinois Indiana Kansas Kentucky Massachusetts Maryland Maine Michigan Minnesota Missouri Mississippi Montana North Carolina North Dakota Nebraska New Hampshire New Jersey New Mexico Nevada New York Ohio Oklahoma Oregon Pennsylvania Tennessee Texas Utah Virginia Washington Wisconsin West Virginia Wyoming
What does Class II mean?
Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.
Source: FDA (medical devices), recall Z-1729-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.