FDA Medical devices recall ·

CELL-DYN Ruby Product Name: CELL-DYN Ruby Model recall

Recall number
Z-1656-2025
Date
(initiated March 18, 2025)
Company / brand
Abbott Laboratories
100 Abbott Park Rd Bldg Ap8b, Abbott Park, IL, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
45
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: CELL-DYN Ruby Product Name: CELL-DYN Ruby Model/Catalog Number: 04U42-84 Software Version: All software versions Product Description: CELL-DYN Ruby

Why was it recalled?

When expired reagents are scanned or manually entered, the system will change the expiration date to current or future date without notifying user that an expired reagent is being used. Use of expired reagents is against the guidelines of the Operator's Manual.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

UDI-DI 00380740017170 SNs: 35939BG 35979BG 36019BG 36103BG 36115BG 36143BG 36171BG 36177BG 36361BG 36788BG 36866BG 54095BG 54431BG 54686BG 54787BG 54900BG 55634BG 55644BG 55656BG 55657BG 55659BG 55662BG 55665BG 55672BG 55673BG 55747BG 55859BG 55879BG 70021BG 70161BG 70204BG 70463BG 70713BG 70931BG 71185BG 71387BG 71388BG 71397BG 71398BG 71476BG 71503BG 71626BG 71637BG 71638BG 71710BG

Distribution

Worldwide distribution: US (nationwide) to the following states: AR, AZ, CA, CO, FL, GA, IA, IL, KS, LA, MA, MD, MI, MN, MO, MS, ND, NE, NJ, NV, NY, OH, OK, PA, PR, SC, TN, TX, UT, VA, WA, WI, WV, WY and OUS (foreign) to countries of: Afghanistan, Albania, Algeria, Angola, Anguilla, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Barbados, Belarus, Belgium, Belize, Bolivia, Bosnia and Herzegovina, Brazil, Brunei, Bulgaria, Canada, Cayman Islands, Chile, China, Colombia, Croatia, Czech Rebublic, Ecuador, Egypt, El Salvador, Ethiopia, France, Gambia, Gaza & Jericho, Germany, Greece, Guatemala, Guana, Haiti, Honduras, Hong King, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Kuwait, Kyrgyzstan, Lebanon, Lithuania, Luxembourg, Malaysia, Maldives, Mali, Malta, Mexico, Moldova, Montenegro, Morocco, Myanmar, New Caledonia, Nicaragua, Nigeria, Norway, Oman, Pakistan, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Romania, Russia, Rwanda, Saudi Arabia, Serbia, Singapore, Slovakia, Slovenia, South Africa, Spain, SRI Lanka, St. Lucia, Taiwan, Tajikistan, Tanzania, Thailand, Trinidad & Tobago, Tunisia, Turk & Caicos, Turkey, UAE, Uganda, UK, Ukraine, Uruguay, Uzbekistan, Vietnam, and Yemen.

States: Arkansas Arizona California Colorado Florida Georgia Iowa Illinois Kansas Louisiana Massachusetts Maryland Michigan Minnesota Missouri Mississippi North Dakota Nebraska New Jersey Nevada New York Ohio Oklahoma Pennsylvania Puerto Rico South Carolina Tennessee Texas Utah Virginia Washington Wisconsin West Virginia Wyoming

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1656-2025. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.