FDA Medical devices recall ·
Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model recall
- Recall number
- Z-1640-2026
- Date
- (initiated February 25, 2026)
- Company / brand
- Olympus Corporation of the Americas
3500 Corporate Pkwy, Center Valley, PA, United States - Category
- Medical devices › Class II
- Classification
- Class II
- Status
- Ongoing
- Units affected
- 444 units
- Hazard tags
- Laceration
- Initiated by
- Voluntary: Firm initiated
Read the official FDA notice ↗
Product
Brand Name: Olympus PK Cutting Forceps, 5mm 33cm Product Name: PK Cutting Forceps Model/Catalog Number: PK-CF0533 Product Description: An assembly of devices intended to generate and apply radio-frequency (RF) alternating current to soft tissues for cutting and coagulation during an endoscopic or open surgical procedure.
Why was it recalled?
Olympus identified that the Cutting Forceps contain components for which the supplier did not adequately validate the welding process. Defective welds can result in the Cutting Forceps' jaw breaking during clinical use.
What should you do?
FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.
Product codes & lot numbers
Model: PK-CF0533; UDI: 00821925035867; Lot Numbers: All unexpired;