FDA Medical devices recall ·

Custom Abutment AS Ti, for Straumann RC, incl recall

Recall number
Z-1633-2026
Date
(initiated March 17, 2026)
Company / brand
Straumann USA LLC
60 Minuteman Rd, Andover, MA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
7 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Custom Abutment AS Ti, for Straumann RC, incl. TAN screw; Catalog Number/Art.: 010.5438;

Why was it recalled?

A material mix-up had occurred during the manufacturing of Straumann¿ Custom Abutments. More specifically, a raw material bar of TAN-material (SAP #701924 / Lot 7130439) was incorrectly identified as a Titanium Bar Ti Gr4 D (SAP #701922 / Lot 7130430). As a result, RC Titanium abutments were manufactured using the TAN Bar material on Mill-from-Bar Machine 396-06.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Catalog Number/Art.: 010.5438; UDI-DI: 07630031773937; Order number (lot): 71315US_00729_01_b00, 91076US_03269_00_b00, 72509US_01048_00_b00, 72839US_00065_00_b00, 39133US_11508_02_b00, 4202CA_00558_00_b00, 72630CA_00109_00_b00;

Distribution

Worldwide - US Nationwide distribution in the states of CA, CO, FL, IA, IL, MN, MO, MT, OR, WA; and the country of Canada.

States: California Colorado Florida Iowa Illinois Minnesota Missouri Montana Oregon Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1633-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.