FDA Medical devices recall ·

Rover Product Name: Mobile X-ray System Model recall

Recall number
Z-1578-2025
Date
(initiated March 28, 2025)
Company / brand
Micro-X Ltd.
Unit 14 6 Mab Eastern Promenade, Tonsley, Australia
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
31 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Rover Product Name: Mobile X-ray System Model/Catalog Number: MXU-RV71 Product Description: Rover Mobile X-rays System ("Rover") is a mobile imaging system that incorporates a self-contained X-ray generator, imaging display and software for acquiring medical diagnostic images outside a standard X-ray room. It is designed specifically for Digital Radiography (DR) with vendor-supported detectors. The Rover is designed to perform radiographic X-ray examinations for disease/injury detection, diagnosis, medical and surgical treatment planning, and therapy monitoring, where patient condition contraindicates transport to a fixed X-ray unit The Rover can be used on pediatric and adult patients, in all patient treatment areas Component: N/A

Why was it recalled?

Potential for early life x-ray tube failure for mobile x-ray system due to excessive thermal load on the anode.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Serial Number/UDI: MSN0334 (01)09357123000051(11)220921(21)00334; MSN0350 (01)09357123000037(11)230328(21)00350 ; MSN0352 (01)09357123000051(11)230823(21)00352; MSN0364 01)09357123000051(11)230705(21)00364; MSN0382 (01)09357123000051(11)230809(21)00382; MSN0383 (01)09357123000051(11)230814(21)00383; MSN0384 (01)09357123000051(11)230816(21)00384; MSN0385 (01)09357123000051(11)230906(21)00385; MSN0386 (01)09357123000051(11)230908(21)00386; MSN0387 (01)09357123000051(11)230914(21)00387; MSN0388 (01)09357123000051(11)230916(21)00388; MSN0389 (01)09357123000051(11)230918(21)00389; MSN0390 (01)09357123000051(11)231013(21)00390; MSN0395 (01)09357123000051(11)231106(21)00395; MSN0397 (01)09357123000051(11)231118(21)00397; MSN0399 (01)09357123000051(11)240122(21)00399; MSN0400 (01)09357123000051(11)240207(21)00400; MSN0401 (01)09357123000051(11)240131(21)00401; MSN0402 01)09357123000051(11)240202(21)00402; MSN0403 (01)09357123000051(11)240205(21)00403; MSN0405 (01)09357123000051(11)240206(21)00405; MSN0408 (01)09357123000051(11)240308(21)00408; MSN0409 (01)09357123000051(11)240312(21)00409; MSN0410 (01)09357123000051(11)240318(21)00410; MSN0411 (01)09357123000051(11)240319(21)00411; MSN0422 (01)09357123000051(11)240724(21)00422; MSN0423 (01)09357123000051(11)240808(21)00423; MSN0424 (01)09357123000051(11)240809(21)00424; MSN0425 (01)09357123000051(11)240812(21)00425; MSN0426 (01)09357123000051(11)240812(21)00426; MSN0427 (01)09357123000051(11)240812(21)00427;

Distribution

US Nationwide distribution in the states of CA, NJ, and Puerto Rico (US territory).

States: California New Jersey Puerto Rico

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.