FDA Medical devices recall ·

STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model recall

Recall number
Z-1491-2026
Date
(initiated February 4, 2026)
Company / brand
Diagnostica Stago, Inc.
5 Century Dr, Parsippany, NJ, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
369
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: STA Liatest Free Protein S Product Name: STA Liatest Free Protein S Model/Catalog Number (REF): 00516 Product Description: The STA¿ - Liatest¿ Free Protein S kits are intended for use on STA-R¿ and STA Compact¿ for the quantitative antigenic assay of free protein S in human citrated plasma by the immuno-turbidimetric method. Component: Not Applicable

Why was it recalled?

The potential of out-of-range results and an underestimation of the free protein S level in normal patient plasmas. The defect only affects normal-range values.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Model/Catalog Number (REF): 00516; UDI: (1) (01)03607450005165(11)241001(17)260131(10)271971(241)00516, (2) (01)03607450005165(11)250221(17)260531(10)272859(241)00516, (3) (01)03607450005165(11)250306(17)260630(10)273018(241)00516, (4) (01)03607450005165(11)250430(17)260831(10)273426(241)00516; Lot numbers: (1) 271 971, (2) 272 859, (3) 273 018, (4) 273 426; Lot Code: 271 971 (01)03607450005165(11)241001(17)260131(10)271971(241)00516

Distribution

Worldwide - US Nationwide distribution in the states of AZ, CA, CO, FL, GA, HI, IL, KY, MI, MN, MO, MT, NC, ND, NJ, NY, OH, PA, TN, TX, UT, VA, WA and the countries of ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BIELORUSSIE, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COSTA RICA, DENMARK, FINLAND, FRANCE, GERMANY, GREECE, GUADELOUPE, GUATEMALA, INDIA, IRAQ, IRELAND, ITALY, JORDAN, KENYA, KOREA REPUBLIC OF, LITUANIA, LUXEMBOURG, MALAYSIA, MEXICO, MOROCCO, NETHERLANDS, NEW CALEDONIA, NEW ZEALAND, PAKISTAN, PALESTINIAN TERRITORY, PARAGUAY, PORTUGAL, ROMANIA, RUSSIE, SAUDI ARABIA, SERBIA, SOUTH AFRICA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TUNISIE, TURKEY, UNITED ARAB EMIRATES, UNITED KINGDOM, URUGUAY;

States: Arizona California Colorado Florida Georgia Hawaii Illinois Kentucky Michigan Minnesota Missouri Montana North Carolina North Dakota New Jersey New York Ohio Pennsylvania Tennessee Texas Utah Virginia Washington

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1491-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.