FDA Medical devices recall ·

Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system… recall

Recall number
Z-1339-2026
Date
(initiated December 12, 2025)
Company / brand
Elekta, Inc.
400 Perimeter Center Ter NE Ste 50, Atlanta, GA, United States
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
N/A
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Leksell GammaPlan Product Name: Radionuclide RT Treatment planning system Software Version: 11.1 Product Description: Leksell GammaPlan¿ is a computer-based system designed for Leksell Gamma Knife¿ treatment planning.

Why was it recalled?

Due to an error when defining a new stereotactic reference that users fail to first change an already defined obsolete reference to a pre-plan reference. This may result in a treatment plan that potentially irradiates the incorrect patient location if the error is not detected by users.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

Lot Code: 07340048311236 (LGP 11.1)

Distribution

Worldwide - U.S. Nationwide distribution including in the states of AK, AZ, CA, CO, ST, FL, GA, HI, IA, IL, LA, MA, MI, MN. MO, NC, NJ, NM, NY, OH, OK, OR, PA, PR, SC, SD, TN, TX, UT, VA, WA, WI, and WV. The countries of Algeria, Argentina, Australia, Belarus, Belgium, Brazil, Bulgaria, Canada, China, Colombia, Croatia, Czech Republic, Dominican Republic, Ecuador, Egypt, El Salvador, France, Georgia, Germany, Hong Kong, India, Iran, Iraq, Italy, Japan, Jordan, Republic of Korea, Kuwait, Libyan Arab Jamahiriya, Lithuania, Malaysia, Mexico, Morocco, Netherlands, Norway, Pakistan, Peru, Poland, Romania, Russian Federation, Saudi Arabia, South Africa, Spain, Sweden, Taiwan, Thailand, Turkey, United Kingdom, and Vietnam.

States: Alaska Arizona California Colorado Florida Georgia Hawaii Iowa Illinois Louisiana Massachusetts Michigan Minnesota Missouri North Carolina New Jersey New Mexico New York Ohio Oklahoma Oregon Pennsylvania Puerto Rico South Carolina South Dakota Tennessee Texas Utah Virginia Washington Wisconsin West Virginia

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1339-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.