Molift Product Name: Molift Mover 205 Model recall

Recall number
Z-1308-2026
Date
(initiated November 17, 2025)
Company / brand
Etac A/S
Parallelvej 3, Gedved, N/A, Denmark
Category
Medical devices › Class II
Classification
Class II
Status
Ongoing
Units affected
95 units
Initiated by
Voluntary: Firm initiated

Read the official FDA notice ↗

Product

Brand Name: Molift Product Name: Molift Mover 205 Model/Catalog Number: (1) M14000, (2) M1401, (3) M1402, (4) M14050, (5) M14150, (6) M1402-01, (7) M14050-00 (8) M14001;

Why was it recalled?

A customer experienced that during assembly and load testing of the device, the bolt for mounting the lifting bar broke. Upon further investigation of this customer complaint, it became clear that the defect occurs as a result of deformation of the materials involved between the lifting bar and the arm of the device. Such deformation can lead to limited mobility in the joint, which prevents proper functionality of the equipment.

What should you do?

FDA enforcement reports do not list a consumer remedy. Check the product codes below against what you have; if it matches, stop using it and follow the instructions in the firm's or FDA's official notice, or contact the company. If you have health concerns, contact your healthcare provider.

Product codes & lot numbers

(1) Model/Catalog Number: M14000; UDI-DI: 05707995032270; Serial Numbers: 616606, 616604, 591003, 591002, 590906, 590901, 590904, 590903, 590902, 591001, 604101, 604005, 591703, 591007, 591006, 591005, 591004, 615105, 615104, 615103, 615102, 615101, 604004, 604003, 604002, 604001, 614202, 614201, 591702, 591701, 616603, 615106, 590905; (2) Model/Catalog Number: M1401; UDI-DI: 05707995032270; Serial Numbers: 593302, 593301, 593303, 612507, 612506, 612508, 612505, 612503, 620702, 612504, 612502, 612501, 593304, 593305, 620703, 620701; (3) Model/Catalog Number: M1402; UDI-DI: 05707995032270; Serial Numbers: 612601; (4) Model/Catalog Number: M14050; UDI-DI: 05707995032270; Serial Numbers: 590801, 596701, 585701, 584501, 591801, 604405, 604406, 604402, 604404, 604403, 604401, 604303, 604302, 604301; (5) Model/Catalog Number: M14150; UDI-DI: 05707995032270; Serial Numbers: 585811, 585810, 585809, 585808, 585806, 585805, 585804, 585803, 585801, 596604, 596603, 596602, 596601, 596611, 596610, 596609, 596608, 596607, 596606, 596605, 585807, 585802; (6) Model/Catalog Number: M1402-01; UDI-DI: 05707995032270; Serial Numbers: 612701, 612703, 612702; (7) Model/Catalog Number: M14050-00; UDI-DI: 05707995032270; Serial Numbers: 584307; (8) Model/Catalog Number: M14001; UDI-DI: 05707995032270; Serial Numbers: 603501, 599801, 599802, 604202, 604201;

Distribution

Worldwide - US Nationwide distribution in the state of PA and the countries of Australia, Czech Republic, Finland, France, Germany, Israel, Italy, Japan, New Zealand, Norway, Poland, Qatar Spain, Sweden, Switzerland, UK.

States: Pennsylvania

What does Class II mean?

Class II means use of or exposure to the product may cause temporary or medically reversible adverse health consequences, or the probability of serious harm is remote.

Source: FDA (medical devices), recall Z-1308-2026. Official notice. Republished from public data; RecallFlag is not affiliated with the FDA.